Reliability of retinal vessel calibre measurements using a retinal oximeter

Abstract

Background

Summarised retinal vessel diameters are linked to systemic vascular pathology. Monochromatic images provide best contrast to measure vessel calibres. However, when obtaining images with a dual wavelength oximeter the red-free image can be extracted as the green channel information only which in turn will reduce the number of photographs taken at a given time. This will reduce patient exposure to the camera flash and could provide sufficient quality images to reliably measure vessel calibres.

Methods

We obtained retinal images of one eye of 45 healthy participants. Central retinal arteriolar and central retinal venular equivalents (CRAE and CRVE, respectively) were measured using semi-automated software from two monochromatic images: one taken with a red-free filter and one extracted from the green channel of a dual wavelength oximetry image.

Results

Participants were aged between 21 and 62 years, all were normotensive (SBP: 115 (12) mmHg; DBP: 72 (10) mmHg) and had normal intra-ocular pressures (12 (3) mmHg). Bland-Altman analysis revealed good agreement of CRAE and CRVE as obtained from both images (mean bias CRAE = 0.88; CRVE = 2.82).

Conclusions

Summarised retinal vessel calibre measurements obtained from oximetry images are in good agreement to those obtained using red-free photographs.

Peer Review reports

Background

Retinal vascular calibres can be measured non-invasively using retinal photographs. In the last 10–15 years a standardised approach to measure and calculate summarised retinal vessel calibres has evolved [1, 2]. Thereby the summarised arteriolar and venular calibres are referred to as central retinal arteriolar equivalent (CRAE) and central retinal venular equivalent (CRVE) from which the arterio-venous ratio (AVR) can be calculated. While the AVR is a poor marker of vascular changes at the retinal level [3, 4], CRAE and CRVE have been shown more clinically useful. In hypertensive patients for example, a decreasing CRAE over time is associated with increased arterial stiffness (in early-stage hypertension) and has shown potential to be predictive in regards to cardiovascular mortality and morbidity. This provides data on the potential of CRAE to become a useful marker for risk stratification in hypertensive (HT) patients [5]. A systematic review by Ding J and colleagues [6] evaluating data of 10229 participants revealed an association of decreased CRAE and increased CRVE with an increased risk of HT. Larger retinal venular calibres are also predictive of stroke [7]. Furthermore, a decrease in CRAE and an increase in CRVE over time are predictive of microvascular outcome in type 1 Diabetes Mellitus (DM) [8].

Recently, there has been an increasing amount of ocular imaging devices available. Each device is developed to assess one or more markers. However, when measuring multiple parameters it is not practical to move participants constantly between instruments. Another limiting factor is acquisition and exposure time which can lead to non-compliance and increased drop-out rates of study participants.

Retinal vessel calibres are not only important to assess in regards to cardiovascular risk and microvascular changes at the retinal level but form an integral part of the analysis for retinal oximetry as it has been shown that the oxygen saturation is depending on vessel diameter [9]. Blondal and colleagues [10] have shown that diameter measurements obtained with their oximeter show good repeatability. While a given marker can be measured by using different techniques, this does not necessarily mean that these techniques can be used interchangeably (i.e. have comparable results), owing to their inherent differences. Here, we set out to explore two such markers – CRAE and CRVE – obtained with two different techniques: one using a red-free filter and the other one using a dual wavelength filter used for taking retinal oximetry images and extracting only the information from the green channel. CRAE and CRVE are more commonly used rather than using individual vessel calibres when examining their associations with vascular pathology. Hence, the focus of this study was not on individual vessel calibre agreement but on summarised vessel calibre agreement when using the two aforementioned acquisition techniques.

Methods

Subjects

The study followed the guidelines of the Declaration of Helsinki and was approved by the Aston Optometry and Audiology Research Ethics Committee. All participants gave written informed consent prior to inclusion in the study. We included 45 arbitrarily selected eyes (25 left and 20 right eyes) of 45 healthy participants (age range: 21 – 62 years). Twelve subjects were of South Asian (SA) background and thirty-three of Caucasian (CA) origin. All participants initially underwent non-contact tonometry (Pulsair, Keeler, UK) followed by pupil dilation using one drop of Tropicamide 1 % (Bausch & Lomb, UK). After a minimum of 15–20 min acclimatisation in a temperature controlled room (21 ± 2 °C) systemic blood pressure and heart rate was measured using a digital sphygmomanometer (UA-767, A&D Instruments Ltd., UK).

Image acquisition and analysis

Following full pupil dilation, first the monochromatic (red-free) and after a resting period of a minimum of 5 min the oximetry (dual-wavelength) retinal photographs were obtained using a Zeiss FF450+ fundus camera (Zeiss, Germany and Imedos Systems, Germany) with the optic nerve head (ONH) centred and the camera field angle set to 30° (see sample image and measurement area in Fig. 1). One image obtained with the dual wavelength filter (specifications: 548 ± 10 nm (full width at half maximum) and 610 ± 10 nm) and one monochromatic image were used to measure the CRAE and CRVE of each image; more detail on the spectral transmission of the dual wavelength filter used for retinal oximetry can be found elsewhere [9]. Room illumination was switched off during imaging.

Fig. 1
figure 1

Sample retinal image with optic nerve head centred and measurement zone from which retinal vessel calibres were obtained

In brief, retinal vessel diameters were measured using a semi-automated piece of software (VesselMap 2, Imedos, Germany) by a single observer (OS). Following image selection, two concentric rings with ½ Disc Diameter (DD) and 1 DD were placed distant from the ONH. The same vessels were assessed in both images. The six largest retinal arteries and veins passing through the ring segment (see Fig. 1) were included in the analysis as recommended by Knudtson and colleagues [2]. The formulae used to derive CRAE and CRVE after three iterations are provided in Eqs. 1, 2 (Parr-Hubbard method)1 below:

$$ {\mathrm{W}}_{\mathrm{artery}}=\sqrt{0.87\ {W}_1^2+1.01\ {W}_2^2-0.22{W}_1{W}_2-10.76} $$
(1)
$$ {\mathrm{W}}_{\mathrm{vein}}=\sqrt{0.72{W}_1^2+0.91{W}_2^2+450.05} $$
(2)

where Wartery denotes an estimated arteriolar trunk width from the pair of the narrower and wider retinal arteriolar branches, W1 and W2, respectively. Similarly, Wvein denotes an estimated venular trunk width from the pair of the narrower and wider retinal venular branches, W1 and W2, respectively. After the first formula iteration the six blood vessels are reduced to three estimated diameters. The narrower and wider vessels of these three are again combined into a new trunk estimate, whereas the remaining vessel is carried over to the next (and final) formula iteration to derive the CRAE or CRVE. From these two parameters AVR is then calculated as follows: AVR = CRAE/ CRVE.

Statistical analysis

Bland-Altman analysis was performed to evaluate the agreement [11, 12] between the calibre parameters obtained (CRAE, CRVE and AVR) from the two images. For ease of visualisation we have included a scatterplot for CRAE and CRVE values for the two sets of images. All data were analysed using GraphPad Prism (Version 6.05, GraphPad Software, California USA).

Results

All forty-five participants (mean (SD) age: 34 (10) yrs) were normotensive (SBP: 115 (12) mmHg; DBP: 72 (10) mmHg) and had IOPs within a normal range (12 (3) mmHg).

Results from the Bland-Altman analysis showed good agreement (mean bias of CRAE, CRVE and AVR is approaching zero) between the two images under comparison. Detailed results including the mean bias (or sometimes referred to as the mean difference), upper and lower limits of agreement can be found in Table 1 and as a graphical representation in the four plots in Fig. 2.

Table 1 Summarised retinal diameters and results from the Band-Altman analysis
Fig. 2
figure 2

Bland Altman plots for (a) CRVE; (b) CRAE and (d) AVR. The solid horizontal line denotes the mean difference between the two summarised vessel calibres as obtained from the red-free and the oximetry image. The two dotted lines show the ±1.96 standard deviations of the difference (denoting the upper and lower limits of agreement). Scatter plot showing the correlation of (c) CRAE (red) and CRVE (blue) between the two images

Discussion

The Bland-Altman analysis showed that the vessel parameters obtained from both images are in good agreement. When observing the plots it is also apparent that the larger arteriolar and venular equivalents show less spread. This is not surprising, as larger arteries and veins can be measured with higher accuracy due to their size and contrast profile.

Retinal vascular calibres can be used as standalone measurements to evaluate the retinal microcirculation, associations with systemic and ocular vascular disease such as glaucoma [13], diabetes [8, 14, 15], hypertension [5, 16], stroke [17] and cardiovascular abnormalities [18, 19]. Since the possibility of obtaining more “dynamic” and “metabolic” data of the ocular circulation through the introduction of retinal vessel analyser (RVA) [20] and retinal oximetry [21], vessel calibres are more commonly used in conjunction with these more novel parameters as well as forming the foundation for correcting influences of vessel size [8]. However, when calculating parameters such as retinal vessel oxygen saturation and blood flow (BF), it is paramount to have accurate vessel calibre measurements as they form the basis of the calculation for some (i.e. BF) as well as complementing other parameters for a more complete insight into the retinal microcirculation.

While some oximeters integrate the vessel calibre measurements into their saturation calculation, they don’t provide automatic output of summarised vessel calibres. These can however be calculated by extracting the green channel information (i.e. red-free image) of the image obtained as in our example. The hitherto additional parameter acquired can be used to compare to previously obtained vessel calibres as well as in conjunction with other metabolic or structural markers measured as part of an examination. The use of the oximeter output rather than an additional monochromatic image using a standard green filter also means less exposure for the patient and shorter acquisition time for the clinician.

Conclusion

Retinal photographs obtained by a retinal oximeter can be used to measure and calculate summarised retinal vessel calibres in good agreement with those obtained using a standard red-free filter. This enables the extraction of another important marker for the assessment of retinal microvasculature in addition to retinal vessel oxygen saturation values without the need of taking additional monochromatic images.

References

  1. Hubbard LD, Brothers RJ, King WN, Clegg LX, Klein R, Cooper LS, et al. Methods for evaluation of retinal microvascular abnormalities associated with hypertension/ sclerosis in the Atherosclerosis Risk In Communities study. Ophthalmology. 1999;106:2269–80.

    Article  CAS  PubMed  Google Scholar 

  2. Knudtson MD, Lee KE, Hubbard LD, Wong TY, Klein R, Klein BE. Revised formulas for summarizing retinal vessel diameters. Curr Eye Res. 2003;27:143–9.

    Article  PubMed  Google Scholar 

  3. Chester EM. The ocular fundus in systemic disease. Chicago: Medical Publishers Inc.; 1977.

    Google Scholar 

  4. Liew G, Sharrett AR, Kronmal R, Klein R, Wong TY, Mitchell P, et al. Measurement of retinal vascular caliber: issues and alternatives to using the arteriole to venule ratio. Invest Ophthalmol Vis Sci. 2007;48(1):52–7.

    Article  PubMed  Google Scholar 

  5. Triantafyllou A, Anyfanti P, Zabulis X, Gkaliagkousi E, Petidis K, Triantafyllou G, et al. Association between retinal vessel caliber and arterial stiffness in a population comprised of normotensive to early-stage hypertensive individuals. Am J Hypertens. 2014;27(12):1472–8.

    Article  PubMed  Google Scholar 

  6. Ding J, Wai KL, McGeechan K, Ikram MK, Kawasaki R, Xie J, et al. Retinal vascular calibre and the development of hypertension: a meta-analysis of individual participant data. J Hypertens. 2014;32(2):207–15.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. McGeechan K, Liew G, Macaskill P, Irwig L, Klein R, Klein BE, et al. Prediction of incident stroke events based on retinal vessel caliber: A systematic review and individual-participant meta-analysis. Am J Epidemiol. 2009;170:1323–32.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Broe R, Rasmussen ML, Frydkjaer-Olsen U, Olsen BS, Mortensen HB, Hodgson L, et al. Retinal vessel calibers predict long-term microvascular complications in type 1 diabetes: the Danish cohort of pediatric diabetes 1987 (DCPD1987). Diabetes. 2014;63:3906–14.

    Article  CAS  PubMed  Google Scholar 

  9. Hammer M, Vilser W, Riemer T, Schweitzer D. Retinal vessel oximetry-calibration, compensation for vessel diameter and fundus pigmentation, and reproducibility. J Biomed Opt. 2008;13(5):054015.

    Article  PubMed  Google Scholar 

  10. Blondal R, Sturludottir MK, Hardarson SH, Halldorsson GH, Stefansson E. Reliability of vessel diameter mesurements with a retinal oximeter. Graefes Arch Clin Exp Ophthalmol. 2001;249:1311–7.

    Article  Google Scholar 

  11. Armstrong RA, Davies LN, Dunne MCM, Gilmartin B. Statistical guidelines for clinical studies of human vision. Ophthalmic Physiol Optics. 2011;31:123–36.

    Article  Google Scholar 

  12. McAlinden C, Khadka J, Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Optics. 2011;31:330–8.

    Article  Google Scholar 

  13. Kawasaki R, Wang JJ, Rochtchina E, Lee AJ, Wong TY, Mitchell P. Retinal vessel caliber is associated with the 10-year incidence of glaucoma: the Blue Mountains Eye Study. Ophthalmology. 2013;120(1):84–90.

    Article  PubMed  Google Scholar 

  14. Kifley A, Wang JJ, Cugati S, Wong TY, Mitchell P. Retinal vascular caliber and the long-term risk of diabetes and impaired fasting glucose: the Blue Mountains Eye Study. Microcirculation. 2008;15(5):373–7.

    Article  CAS  PubMed  Google Scholar 

  15. Nguyen TT, Wang JJ, Sharrett AR, Islam FM, Klein R, Klein BE, et al. Relationship of retinal vascular caliber with diabetes and retinopathy: the Multi-Ethnic Study of Atherosclerosis (MESA). Diabetes Care. 2008;31(3):544–9.

    Article  CAS  PubMed  Google Scholar 

  16. Wang JJ, Rochtchina E, Liew G, Tan AG, Wong TY, Leeder SR, et al. The long-term relation among retinal arteriolar narrowing, blood pressure, and incident severe hypertension. Am J Epidemiol. 2008;168(1):80–8.

    Article  PubMed  Google Scholar 

  17. Wieberdink RG, Ikram MK, Koudstaal PJ, Hofman A, Vingerling JR, Breteler MM. Retinal vascular calibers and the risk of intracerebral hemorrhage and cerebral infarction: the Rotterdam Study. Stroke. 2010;41(12):2757–61.

    Article  PubMed  Google Scholar 

  18. Josef P, Ali I, Ariel P, Alon M, Nimer A. Relationship between retinal vascular caliber and coronary artery disease in patients with non-alcoholic fatty liver disease (NAFLD). Int J Environ Res Public Health. 2013;10(8):3409–23.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Miller RG, Prince CT, Klein R, Orchard TJ. Retinal vessel diameter and the incidence of coronary artery disease in type 1 diabetes. Am J Ophthalmol. 2009;147(4):653–60.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Vilser W, Nagel E, Lanzl I. Retinal vessel analysis--new possibilities. Biomed Tech (Berl). 2002;47(Suppl 1 Pt 2):682–5.

    Article  Google Scholar 

  21. Hardarson SH. Retinal oximetry. Acta Ophthalmol. 2013;91(Thesis 2):1–47.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We would like to thank Dr Richard Armstrong for his invaluable advice on statistical analysis and acknowledge Omar Soomah for his help with image processing.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Rebekka Heitmar.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

RH and AAK have collected and analysed the data. RH and AAK have written the manuscript together. Both authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Heitmar, R., Kalitzeos, A.A. Reliability of retinal vessel calibre measurements using a retinal oximeter. BMC Ophthalmol 15, 184 (2015). https://doi.org/10.1186/s12886-015-0174-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-015-0174-0

Keywords