Visualization of precut DSAEK and pre-stripped DMEK donor corneas by intraoperative optical coherence tomography using the RESCAN 700

Abstract

Background

To report the feasibility of intraoperative spectral domain optical coherence tomography (OCT) using the RESCAN 700 for the visualization and evaluation of precut Descemet’s stripping automated endothelial keratoplasty (DSAEK) and prestripped Descemet’s membrane endothelial keratoplasty (DMEK) donor tissue.

Methods

Precut DSAEK (n = 11), prestripped DMEK (n = 4) preserved in a viewing chamber were examined by intraoperative OCT. Wholly intact donor corneas for penetrating keratoplasty (PK) (n = 8) were also examined as controls. The obtained images were analyzed qualitatively for characteristics of each type of donor.

Results

For each type of donor, characteristic images were consistently obtained by intraoperative OCT in both the front and back views through the viewing chamber. In wholly intact donors for PK, appearance of normal corneal curvature and stromal texture with high reflectivity of epithelium and endothelium cell layers were clearly visualized. In precut DSAEK donors, precut lines were characteristically visualized in addition to the intact donor cornea images. In prestripped DMEK donors, identical OCT images to the intact donor cornea were noted when observed from the anterior surface. However, peripheral partial detachments of Descemet’s membrane were characteristically observed in all prestripped DMEK donors when viewed from the back of the viewing chamber.

Conclusion

Rapid visualization and rough evaluation of donor tissues for PK, precut DSAEK and prestripped DMEK donor corneas by intraoperative OCT was consistently possible through the viewing chamber. Therefore, this device may be used as an alternative of AS-OCT when the eyebank does not have their own AS-OCT. Although the peripheral detachment in DMEK donors are quite common and clinically non-problematic in DMEK donor quality and subsequent DMEK surgeries, it may be useful to distinguish between wholly intact PK donors and prestripped DMEK donors, enabling to prevent mix-ups of donors, especially when several different types of keratoplasties are scheduled in a same day in one operating theater.

Peer Review reports

Background

Descemet’s stripping automated endothelial keratoplasty (DSAEK) [17] or Descemet’s membrane endothelial keratoplasty (DMEK) has been preferably used for the treatment of corneal endothelial dysfunction [817]. Recent statistics showed that these techniques are the most common keratoplasty procedures performed in the U.S. [18]. The use of precut DSAEK tissue [19] or prestripped DMEK tissue [20] supplied by the eyebank has greatly shortened the operation time, avoids a large capital investment in an expensive microkeratome machine for DSAEK and avoids tissue loss due to unsuccessful Descemet’s membrane stripping/harvesting in DMEK. Evaluation of these premade endothelial keratoplasty tissues before surgery can be successfully and effectively performed by slit lamp biomicroscopy and specular microscopy. However, it is difficult to precisely evaluate the depth and quality of the lamellar cut or prestripped status by these modalities.

Previously, anterior segment optical coherence tomography (AS-OCT) has proven useful to screen donor corneas for previous refractive surgeries such as laser in situ keratomileusis [21, 22]. Recently, some eyebanks have adopted AS-OCT as their standard procedures for tissue evaluation [23].

Herein, we show the feasibilities of intraoperative spectral domain OCT using the RESCAN 700 (Carl Zeiss Meditec, Germany) [24, 25] for the rapid visualization and rough evaluation of precut DSAEK and prestripped DMEK donor tissue.

Methods

The study was approved by the Ethical Committee of Kanazawa University Graduate School of Medical Science and followed the tenets of the Declaration of Helsinki. Internationally shipped precut DSAEK (n = 11) and prestripped DMEK (n = 4) donor corneas from a US eyebank (SightLife™, Seattle, USA) preserved in a viewing chamber were used. For a control, domestic wholly intact donor corneas taken post-mortem for penetrating keratoplasty (PK) (n = 8) were used. Consent was received for the use of these corneas for this research by the donors’ next of kin. The donor cornea for DSAEK was dissected with a microkeratome (ALTK Cbm; Moria, France) equipped with a 300 μm head at the US eyebank. The DMEK donor graft was also prepared at the US eyebank using the submerged cornea using the backgrounds away (SCUBA) technique [15]. All donor tissues (PK tissue, precut DSAEK tissue, pre-stripped DMEK tissue) were kept in Optisol GS solution (Bausch & Lomb Surgical, USA) and internationally shipped by airplane at 4° Celsius. After warming the donor tissues, they were subjected to routine slit lamp biomicroscopy and intraoperative SD-OCT without taking them out of the preservation medium and container.

Intraoperative OCT

The Rescan 700 is a real-time intraoperative spectral domain OCT integrated into the OPMI Lumera 700 microscope (Carl Zeiss Meditec, Germany). The Rescan 700 scan rate was 27,000 axial scans per second. The axial and transverse resolution in tissue are 5 μm and 15 μm, respectively. Each donor tissue was scanned through the transparent window of the corneal viewing chamber, which was placed on a flat table. The OCT images are recorded in a horizontal and vertical orientation. The scan depth was 2 mm and the scan length was adjusted to each donor cornea tissue between 9 and 11 mm (adjustable between 3–16 mm). The surgeon observed the live, uninterrupted intraoperative OCT images during surgery through a small window adjacent to the operating field. Acquired images were evaluated qualitatively for the shape and reflectivity of the donor cornea tissue.

Results

For each type of donor (PK, DSAEK, DMEK donor), characteristic images were consistently and easily obtained by intraoperative OCT in both the front and back views through the viewing chamber.

In wholly intact PK donors, appearance of normal corneal curvature, although not quantitatively measured, and normal stromal texture with high reflectivity of the epithelium and endothelium cell layers were clearly visualized in all donors (n = 8) (Fig. 1).

Fig. 1
figure 1

Intraoperative spectral domain optical coherence tomography (SD-OCT) image of a donor cornea for penetrating keratoplasty observed through a viewing chamber. a Representative microscope image (surgeon’s view) observed from the anterior (left). Normal corneal curvature and stromal texture with high reflectivity of epithelium and endothelium cell layers were observed by intraoperative OCT (right). b Microscopic image observed from the posterior cornea (left) and intraoperative OCT image (right)

In precut DSAEK donors, precut lines were characteristically visualized rapidly in addition to the PK donor cornea images (Fig. 2). Some precut DSEAK donors showed epithelial defect and/or debris on the surface of endothelium; however, these changes usually do not have any clinical significance.

Fig. 2
figure 2

Intraoperative SD-OCT image of a donor cornea for precut Descemet’s stripping automated endothelial keratoplasty (DSAEK) observed through a viewing chamber. a Representative microscope image (surgeon’s view) observed from the anterior (left). Normal corneal curvature and stromal texture together with high reflectivity of the epithelium, microkeratome cut line (arrows), endothelium cell layer and epithelial defect (arrowheads) were observed by intraoperative OCT (right). b Microscopic image observed from the posterior cornea (left) and intraoperative OCT image (right). Characteristic microkeratome cut line (arrows) together with debris (arrowheads) observed on the endothelium cell surface was observed (left)

In prestripped DMEK donors, no characteristic images were noted by SD-OCT when observed from the anterior surface, since peripheral image could not be obtained from the anterior surface observation. However, peripheral partial detachments of Descemet’s membrane were characteristically observed in all prestripped DMEK donors when viewed only from the back of the viewing chamber (Fig. 3). Debris on the surface of endothelium was also seen in some donors.

Fig. 3
figure 3

Intraoperative SD-OCT image of a donor cornea for prestripped Descemet’s membrane endothelial keratoplasty (DMEK) observed through a viewing chamber. a Representative microscope image (surgeon’s view) observed from the anterior (left). Normal corneal curvature and stromal texture were observed by intraoperative OCT (right). It was not possible to observe the prestripping line. b Microscopic image observed from the posterior cornea (left) and intraoperative OCT image (right). Characteristic peripheral partial detachment of Descemet’s membrane was observed (arrows). c Microscopic image of a different area observed from the posterior cornea (left) and intraoperative OCT image (right). Characteristic peripheral partial detachment of Descemetcpes membrane was also noted (arrows)

All donor corneas were transplanted (8 PK, 11 DSAEK, 4 DMEK) to patients, and all corneas remain transparent for at least 12 months.

Discussion

The microscope integrated intraoperative OCT is a relatively new technology. It enables in vivo 3-D visualization of the steps of retina [2628] and cornea [2934] surgeries. The RESCAN 700 is the first commercially available intraoperative OCT system combining a surgical microscope (OPMI LUMERA 700, Carl Zeiss Meditec, Germany) and a spectral domain OCT having a wavelength of 840 nm and a scanning speed of 27,000 A-scans per second. The surgeon does not need to look up from the microscope to see the OCT data since the RESCAN 700 allows the surgeon to simultaneously see the surgical field in both a planar view using the surgical microscope and a cross-sectional view by OCT. Most importantly, images are viewed in real time. The OCT system can be controlled from the microscope’s foot pedal so the surgeon can take videos, snapshots and 3-D OCT images without looking up or stopping the surgery. The usefulness of the RESCAN 700 has been reported for anterior segment surgeries (corneal incisions, scleral closure phacoemulsification groove depth and intraocular lens position) [24], riboflavin penetration in collagen crosslinking [25] and retinal surgeries (evaluation of hyaloid release with triamcinolone and completeness of peel in macular hole and epiretinal membrane) [24]. Recently, we have also confirmed its clinical usefulness in DSAEK for detecting residual fluid between the donor and host cornea which enables a surgeon to decide the appropriate timing to stop corneal massage and push out the residual interface fluid. Also, it is quite useful in DMEK for preventing upside-down DMEK donors in the anterior chamber (data not shown).

Herein, we report the feasibilities of the visualization of donor corneas for PK, precut DSAEK and pre-stripped DMEK by intraoperative OCT using the RESCAN 700. As a result, we showed characteristic images for each type of donor. In wholly intact PK donors, appearance of normal corneal curvature and stromal texture with high reflectivity of the epithelium and endothelium cell layers were clearly visualized. However, the precise corneal curvature could not be measured by RESCAN 700. The images were consistent with those obtained by conventional AS-OCT. In precut DSAEK donors, precut lines were characteristically visualized and the results were also in accordance with those observed by conventional AS-OCT. In prestripped DMEK donors, identical OCT images as the PK donor cornea were noted when observed from the anterior surface. However, peripheral partial detachments of Descemet’s membrane were characteristically observed in all prestripped DMEK donors when viewed from the back of the viewing chamber. Although the peripheral detachment in DMEK donors are quite common and clinically non-problematic in DMEK donor quality and subsequent DMEK surgeries, it may be useful to distinguish between wholly intact PK donors and prestripped DMEK donors by detecting the peripheral detachment sign; it may be difficult or impossible to check the subtle Descemt’s membrane change using slit lamp examination alone.

AS-OCT has been used as a routine procedure for tissue evaluation in some advanced eyebanks to detect previous refractive surgeries and corneal scarring [23]. On the other hand, most eyebanks do not have their own AS-OCT system yet in Japan. Therefore, intraoperative OCT may be used as an alternative device of AS-OCT; it may be able to detect not only corneal donor status, but also corneal scars and previous refractive surgeries just like AS-OCT equipped in eyebank. Another potential usefulness of this device is to routinely check the donor status just before surgery by surgeon to prevent mix-ups of donors, especially when several different types of keratoplasties are scheduled in a same day in one operating theater.

One drawback of this study is that quantitative analysis of the donor status is lacking; the quantitative information such as donor thickness or curvature will definitely strengthen this paper. However, RESCAN 700 currently does not have any quantitative capacity. Development of new software in RESCAN 700 that enables quantitative analysis will enhance the capacity of the device.

Conclusion

In conclusion, rapid visualization and rough evaluation of donor tissues for PK, precut DSAEK and prestripped DMEK donor corneas by intraoperative OCT was consistently possible through the viewing chamber. Although clinically acceptable, peripheral detachment of Descemet’s membrane of prestripped DMEK donors was successfully and characteristically observed by intraoperative OCT. It may be possible to prevent mix-ups of donors by checking those donors just before surgery by surgeon especially several different types of keratoplasties are scheduled in a same day in one operating theater.

Abbreviations

OCT:

Optical coherence tomography

DSAEK:

Descemet’s stripping automated endothelial keratoplasty

DMEK:

Descemet’s membrane endothelial keratoplasty

PK:

Penetrating keratoplasty

References

  1. Melles GR, Eggink FA, Lander F, et al. A surgical technique for posterior lamellar keratoplasty. Cornea. 1998;17:618–26.

    Article  CAS  PubMed  Google Scholar 

  2. Terry MA, Ousley PJ. Small-incision deep lamellar endothelial keratoplasty (DLEK): six-month results in the first prospective clinical study. Cornea. 2005;24:59–65.

    Article  PubMed  Google Scholar 

  3. Terry MA, Ousley PJ. Deep lamellar endothelial keratoplasty: visual acuity, astigmatism, and endothelial survival in a large prospective series. Ophthalmology. 2005;112:1541–8.

    Article  PubMed  Google Scholar 

  4. Price MO, Price Jr FW. Descemet’s stripping with endothelial keratoplasty: comparative outcomes with microkeratome-dissected and manually dissected donor tissue. Ophthalmology. 2006;113:1936–42.

    Article  PubMed  Google Scholar 

  5. Price Jr FW, Price MO. Descemet’s stripping with endothelial keratoplasty in 200 eyes: early challenges and techniques to enhance donor adherence. J Cataract Refract Surg. 2006;32:411–8.

    Article  PubMed  Google Scholar 

  6. Gorovoy MS. Descemet-stripping automated endothelial keratoplasty. Cornea. 2006;25:886–9.

    Article  PubMed  Google Scholar 

  7. Kobayashi A, Yokogawa H, Sugiyama K. Descemet stripping with automated endothelial keratoplasty for bullous keratopathies secondary to argon laser iridotomy--preliminary results and usefulness of double-glide donor insertion technique. Cornea. 2008;27(suppl):S62–9.

    Article  PubMed  Google Scholar 

  8. Melles GR, Lander F, Rietveld FJ. Transplantation of Descemet’s membrane carrying viable endothelium through a small scleral incision. Cornea. 2002;21:415–8.

    Article  PubMed  Google Scholar 

  9. Melles GR, Ong TS, Ververs B, van der Wees J. Descemet’s membrane endothelial keratoplasty (DMEK). Cornea. 2006;25:987–90.

    Article  PubMed  Google Scholar 

  10. Melles GR. Posterior lamellar keratoplasty: DLEK to DSEK to DMEK. Cornea. 2006;25:879–81.

    Article  PubMed  Google Scholar 

  11. Melles GR, Ong TS, Ververs B, van der Wees J. Preliminary clinical results of Descemet membrane endothelial keratoplasty. Am J Ophthalmol. 2008;145:222–7.

    Article  PubMed  Google Scholar 

  12. Ham L, Dapena I, van Luijk C, et al. Descemet membrane endothelial keratoplasty (DMEK) for Fuchs endothelial dystrophy: review of the first 50 consecutive cases. Eye (Lond). 2009;23:1990–8.

    Article  CAS  Google Scholar 

  13. Ham L, Balachandran C, Verschoor CA, et al. Visual rehabilitation rate after isolated Descemet membrane transplantation: Descemet membrane endothelial keratoplasty. Arch Ophthalmol. 2009;127:252–5.

    Article  PubMed  Google Scholar 

  14. Dapena I, Moutsouris K, Droutsas K, et al. Standardized “no-touch” technique for Descemet membrane endothelial keratoplasty. Arch Ophthalmol. 2011;129:88–94.

    Article  PubMed  Google Scholar 

  15. Price MO, Giebel AW, Fairchild KM, et al. Descemet’s membrane endothelial keratoplasty; prospective multi-center study of visual and refractive outcomes and endothelial survival. Ophthalmology. 2009;116:2361–8.

    Article  PubMed  Google Scholar 

  16. Kobayashi A, Yokogawa H, Yamazaki N, et al. In vivo laser confocal microscopy after Descemet’s membrane endothelial keratoplasty. Ophthalmology. 2013;120:923–30.

    Article  PubMed  Google Scholar 

  17. Kobayashi A, Yokogawa H, Yamazaki N, et al. The use of endoillumination probe-assisted Descemet membrane endothelial keratoplasty for bullous keratopathy secondary to argon laser iridotomy. Clin Ophthalmol. 2015;9:91–3.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Eye Bank Association of America. 2014 Eye Banking Statistical Report. Available at www.restoresight.org. Accessed 3 Aug 2016.

  19. Terry MA, Shamie N, Chen ES, Phillips PM, Hoar KL, Friend DJ. Precut tissue for Descemet’s stripping automated endothelial keratoplasty: vision, astigmatism, and endothelial survival. Ophthalmology. 2009;116(2):248–56.

    Article  PubMed  Google Scholar 

  20. Kobayashi A, Murata N, Yokogawa H, Yamazaki N, Masaki T, Sugiyama K. Evaluation of internationally shipped prestripped donor tissue for descemet membrane endothelial keratoplasty by vital dye staining. Cornea. 2015;34(2):225–7.

    Article  PubMed  Google Scholar 

  21. Lin RC, Li Y, Tang M, McLain M, Rollins AM, Izatt JA, Huang D. Screening for previous refractive surgery in eye bank corneas by using optical coherence tomography. Cornea. 2007;26(5):594–9.

    PubMed  Google Scholar 

  22. Priglinger SG, Neubauer AS, May CA, Alge CS, Wolf AH, Mueller A, Ludwig K, Kampik A, Welge-Luessen U. Optical coherence tomography for the detection of laser in situ keratomileusis in donor corneas. Cornea. 2003;22(1):46–50.

    Article  PubMed  Google Scholar 

  23. Bald MR, Stoeger C, Galloway J, Tang M, Holiman J, Huang D. Use of fourier-domain optical coherence tomography to evaluate anterior stromal opacities in donor corneas. J Ophthalmol. 2013;2013:397680.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Ehlers JP, Kaiser PK, Srivastava SK. Intraoperative optical coherence tomography using the RESCAN 700: preliminary results from the DISCOVER study. Br J Ophthalmol. 2014;98(10):1329–32.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Pahuja N, Shetty R, Jayadev C, Nuijts R, Hedge B, Arora V. Intraoperative Optical Coherence Tomography Using the RESCAN 700: Preliminary Results in Collagen Crosslinking. Biomed Res Int. 2015;2015:572698.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Ray R, Baranano DE, Fortun JA, et al. Intraoperative microscope-mounted spectral domain optical coherence tomography for evaluation of retinal anatomy during macular surgery. Ophthalmology. 2011;118:2212–7.

    Article  PubMed  Google Scholar 

  27. Binder S, Falkner-Radler CI, Hauger C, et al. Feasibility of intrasurgical spectral-domain optical coherence tomography. Retina. 2011;31:1332–6.

    Article  PubMed  Google Scholar 

  28. Ehlers JP, Tao YK, Farsiu S, et al. Integration of a spectral domain optical coherence tomography system into a surgical microscope for intraoperative imaging. Invest Ophthalmol Vis Sci. 2011;52:3153–9.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Lim LS, Aung HT, Aung T, Tan DT. Corneal imaging with anterior segment optical coherence tomography for lamellar keratoplasty procedures. Am J Ophthalmol. 2008;145(1):81–90.

    Article  PubMed  Google Scholar 

  30. Ide T, Wang J, Tao A, Leng T, Kymionis GD, O’Brien TP, Yoo SH. Intraoperative use of three-dimensional spectral-domain optical coherence tomography. Ophthalmic Surg Lasers Imaging. 2010;41(2):250–4.

    Article  PubMed  Google Scholar 

  31. Knecht PB, Kaufmann C, Menke MN, Watson SL, Bosch MM. Use of intraoperative fourier-domain anterior segment optical coherence tomography during descemet stripping endothelial keratoplasty. Am J Ophthalmol. 2010;150(3):360–5.

    Article  PubMed  Google Scholar 

  32. Wylegala E, Nowinska AK, Wroblewska-Czajka E, Janiszewska D. Donor disc attachment assessment with intraoperative spectral optical coherence tomography during descemet stripping automated endothelial keratoplasty. Indian J Ophthalmol. 2013;61(9):511–3.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Juthani VV, Goshe JM, Srivastava SK, Ehlers JP. Association between transient interface fluid on intraoperative OCT and textural interface opacity after DSAEK surgery in the PIONEER study. Cornea. 2014;33(9):887–92.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Miyakoshi A, Ozaki H, Otsuka M, Hayashi A. Efficacy of Intraoperative Anterior Segment Optical Coherence Tomography during Descemet’s Stripping Automated Endothelial Keratoplasty. ISRN Ophthalmol. 2014;2014:562062.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgments

None.

Funding

This study was supported by a grant from a Grant-in-Aid for Scientific Research (C) KAKENHI, Japan (16 K11261). The sponsor or funding organization had no role in the design or conduct of this research.

Availability of data and materials

Data sharing not applicable to this article as no datasets were generated or analysed during the current study.

Authors’ contributions

AK, HY, NM carried out the experiment and AK and KS participated in the design of the study. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval and consent to participate

This study was approved by the Ethical Committee of Kanazawa University Graduate School of Medical Science and followed the tenets of the Declaration of Helsinki.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Akira Kobayashi.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kobayashi, A., Yokogawa, H., Mori, N. et al. Visualization of precut DSAEK and pre-stripped DMEK donor corneas by intraoperative optical coherence tomography using the RESCAN 700. BMC Ophthalmol 16, 135 (2016). https://doi.org/10.1186/s12886-016-0308-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-016-0308-z

Keywords