Effects of diurnal, lighting, and angle-of-incidence variation on anterior segment optical coherence tomography (AS-OCT) angle metrics

Abstract

Background

First reported study to assess the effect of diurnal variation on anterior chamber angle measurements, as well as, to re-test the effects of lighting and angle-of-incidence variation on anterior chamber angle (ACA) measurements acquired by time-domain anterior segment optical coherence tomography (AS-OCT).

Methods

A total of 30 eyes from 15 healthy, normal subjects underwent anterior chamber imaging using a Visante time-domain AS-OCT according to an IRB-approved protocol. For each eye, the inferior angle was imaged twice in the morning (8 am – 10 am) and then again in the afternoon (3 pm – 5 pm), under light meter-controlled conditions with ambient room lighting ‘ON’ and lights ‘OFF’, and at 5° angle of incidence increments. The ACA metrics measured for each eye were: angle opening distance (AOD, measured 500 and 750 μm anterior from scleral spur), the trabecular-iris-space area (TISA, measured 500 and 750 μm anterior from scleral spur), and scleral spur angle. Measurements were performed by masked, certified Reading Center graders using the Visante’s Internal Measurement Tool. Differences in measurements between morning and afternoon, lighting variations, and angle of incidence were compared.

Results

Mean age of the participants was 31.2 years (range 23–58). Anterior chamber angle metrics did not differ significantly from morning to afternoon imaging, or when the angle of incidence was offset by 5° in either direction away from the inferior angle 6 o’clock position. (p-value 0.13-0.93). Angle metrics at the inferior corneal limbus, 6 o’clock position (IC270), with room lighting ‘OFF’, showed a significant decrease (p < 0.05) compared to room lighting ‘ON’.

Conclusions

There does not appear to be significant diurnal variation in AS-OCT parameters in normal individuals, but lighting conditions need to be strictly controlled since variation in lighting led to significant variability in AS-OCT parameters. No changes in ACA parameters were noted by varying the angle-of-incidence, which gives confidence in being able to perform longitudinal studies in approximately the same area (plus/minus 5° of original scan location).

Peer Review reports

Background

Primary angle closure glaucoma (PACG) is one of the leading causes of irreversible visual morbidity in the world [1]. The mechanisms are largely connected to poor filtration of aqueous fluid through the anterior chamber angle (ACA) [2]. Evaluation and measurement of the ACA is relevant to risk assessment of angle closure [3]. Gonioscopy, ultrasound biomicroscopy (UBM) and the scheimpflug systems have been used for the visualization and measurement of the ACA [26]. Anterior segment optical coherence tomography (AS-OCT) is a newer non-contact method of imaging which affords the evaluation of the ACA [4, 5]. Certain types of AS-OCT machines, like Visante OCT (Carl Zeiss, Meditec, Dublin, CA, USA), utilize time-domain technology [6, 7], and. allow visualization and evaluation of irido-corneal boundaries and the angle configurations [8, 9]. OCT imaging of the anterior segment could advance detection of angle metrics and reveal anatomical insights into the pathophysiology of angle closure glaucoma [10]. The ACA measurements obtained by AS-OCT have helped much to elucidate the mechanisms of angle closure glaucoma in the management of patients in clinical practice [7, 11, 12].

Low light conditions have been known to be associated with acute attacks of angle closure. Patients with shallow anterior chamber, shorter axial length, and larger lens volume, are statistically demonstrated to have increased risk of angle closure. However, the majority of eyes with these anatomic conditions do not develop angle closure glaucoma [13]. Therefore recent studies have tried to investigate the role of dynamic factors, such as pupil dilation on angle measurements [1416]. Furthermore intraocular pressure, as well as biometric variables, including axial length, central cornea thickness and anterior chamber depth and volume have been shown to demonstrate a diurnal flunctuation [17, 18]. Therefore, it is important to know whether the ACA also demonstrates a change throughout the day. Finally, previous research has also shown marked differences between the measurements of ACA in different quadrants of the eye [19, 20]. Based on these studies, one could hypothesize that there may be differences when the scan line is varied in the same quadrant. This study explored the effect of varying the scan line in the inferior quadrant near the 6 o’clock position of inferior corneal limbus (IC 270°) during the morning and afternoon. We previously demonstrated that Visante Time Domain (TD)-OCT has been repeatable and reproducible for measurement of the angle, allowing us to measure the ACA quantitatively and derive reasonable angle metrics [21].

The objectives of our present study was to first evaluate the changes in ACA metrics for a diurnal effect in asymptomatic normal eyes, second was to examine how the angle of incidence affects measurements, and third was to test the effect of physiologic dilation via room lighting on ACA metrics.

Methods

Fifteen healthy volunteers were recruited for this study which followed the tenets of the Declaration of Helsinki, the Health Insurance Portability and Accountability Act and had the Institutional Review Board approval.

Each subject had both eyes (total of 30 eyes) imaged with the Visante TD-OCT (Carl Zeiss Meditec, Dublin, CA). The first set of images was obtained in the morning under standard room lighting at the IC270 position and also in a darkened room at the IC275, 270, and 265 positions (Fig. 1).

Fig. 1
figure 1

The set of images obtained with time-domain anterior segment optical coherence tomography at the IC275 (a), 270 (b), and 265 (c) positions

The light levels was read at the eye/camera interface with a Sper Scientific Light Meter (FC – 840021, Sper Scientific, Scottsdale, AZ). Each eye was let to adjust to the light level for 1 min in the room before imaging. Two different light levels were used; a light meter reading of 1.0 fc (fc;1 fc 10.77 cd/m2) (lights ON) and .0.0 fc (Lights OFF).

The varying angle of incidence was based on the visualization of the scan line by the OCT technician taking the images. The second set of images was obtained in the afternoon under the same conditions as the morning. The inferior angle opening distance (AOD, measured 500 and 750 μm anterior from Scleral Spur), trabecular-iris space area (TISA, measured 500 and 750 μm anterior from Scleral Spur), and scleral spur angle were computed for each eye and condition.

The Visante TD-OCT internal angle measurement tools were used to determine angle metrics based on the visualization of the scleral spur. All grading of angle metrics was performed by anterior segment OCT graders (HA, KM) at the Doheny Image Reading Center (DIRC). Repeated measures ANOVA and the paired t-test were used. Statistical significance was assumed for p ≤ 0.05 (version 18.0; SPSS Inc., Armonk, NY).

Results

The study evaluated fifteen (10 female and 5 male) subjects. The mean age was 31.2 years (range 23–58). The mean values from 30 eyes for the inferior angle metrics 6 o’clock position (IC270), under two different light conditions in the morning and in the afternoon using the Visante TD-OCT are shown in Tables 1 and 2 (Additional file 1; anterior chamber angle metrics). There were statistically significant differences in AOD500 between the condition of pupillary constriction to room light and physiologic mydriasis both in the morning and in the afternoon (p < 0.001 for both). AOD750 also showed significant difference for both light conditions in the morning and afternoon (p < 0.001 and p = 0.031 respectively). Analogously, TISA500 measured under pupillary constriction to room light showed statistically significant difference compared to the condition under physiologic mydriasis in the morning and in the afternoon respectively (p < 0.001 and p = 0.002). Compared to pupil constriction to light, the mean scleral spur angle increased 9.5 and 12.5% under physiologic mydriasis in the morning and in the afternoon respectively (p < 0.001 for both).

Table 1 Inferior angle anterior chamber angle metrics under two lighting conditions; with the room lights on and off, from two imaging sessions; morning and afternoon imaging, using the TDOCT
Table 2 Angle parameters from the morning imaging session, with the room lights off, with the angle of incidence at 270° and 5° apart in either direction (265° and 275°)

The angle of incidence analysis was explored with the lights off in the morning and in the afternoon. The measurements with the angle of incidence offset by 5° toward the nasal and temporal quadrants (IC265-IC275) and at the 6 o’clock position are shown in Table 2. AOD500 and TISA 500 did not show significant differences at IC265 and IC275 compared to the value at IC270 in the morning (p = 0.8 and p = 0.81 respectively). TISA500 decreased by 3.6 and 4.4% at IC265 and IC275 compared to IC270 in the morning. AOD500 increased by 3.8% at IC265 and decreased by 5.8% at IC275 compared to IC270 in the afternoon respectively but none of these reached the level of significance. AOD750 and TISA 750 did not show significant differences at IC265 and IC275 compared to the value at IC270 in the morning (p = 0.6 and p = 0.86 respectively).

Finally we assessed the variability of the angle parameters in the morning and in the afternoon at 6 o’clock position (Table 3). There was no statistically significant difference between morning and afternoon measurements for the angle parameters at IC270 (P value 0.13-0.93).

Table 3 Angle parameters in the morning and afternoon under low light conditions at IC270

Discussion

The current study evaluated the diurnal, lighting, and angle of incidence variation of inferior angle metrics in phakic healthy volunteers using high resolution cross-sectional images of the anterior segment angle with the Visante TD-OCT. Anterior chamber angle metrics did not differ significantly from morning to afternoon imaging, or when the angle of incidence was offset by 5° in either direction away from the inferior angle 6 o’clock position. However, ACA metrics changed when the illumination at the level of the eye changed. Our results support that lightning conditions should be considered when angle imaging is performed, especially with anterior segment OCT, since the illumination conditions can more accurately and easily be controlled.

Physiologic conditions causing dynamic changes in intraocular structures were evaluated before, particularly the response of the iris to light and dark [22]. Quigley et al., using AS-OCT, found a reduction in the iris cross-sectional area after the pupil dilation and they argued that this could be a potential risk factor for angle closure [23]. They hypothesized that eyes with angle closure lose less iris volume on dilation compared to healthy eyes, contributing to iridotrabecular apposition. Additionally in a different study the iris volume was demonstrated to increase after pupil dilation in narrow-angle eyes predisposing to acute angle closure, whereas it was found to decrease in healthy open-angle eyes [24]. In a previous study, we compared the effects of physiologic versus pharmacologic pupil dilation on ACA measurements obtained with spectral domain optical coherence tomography [14]. This study showed a decrease in inferior angle ACA metrics with physiologic pupil dilation under low light conditions and an increase after pharmacologic pupil dilation in normal eyes. These results are in agreement with findings from this study, suggesting that pupil dilation, that happens in dim lighting leads to narrowing of ACA parameters. Therefore, standardizing of lighting conditions is important for objective measurement of ACA metrics.

Studies have already shown that axial length, intraocular pressure and anterior eye biometrics manifest a diurnal rhythm [17, 25, 26]. However, it is not so well understood whether the ACA also shows nyctohemeral changes. In the present study, we compared OCT scan of the ACA between morning and afternoon and found no significant difference.

In addition, results from this paper do not support a statistically significant difference in ACA metrics in open angle eyes when comparing inferior angle scans with differences in the angle of incidence up to 10° apart. This corroborates findings from a previous study, using both time-domain and spectral-domain OCT and might indicate that regional irregularities of the iris profile usually do not play a role in these small variations of the scan position [27]. However, this finding remains to be proven in narrow angles.

Our study is not without limitations. First, we used 30 normal eyes, which is a relatively small population. However, the findings were highly consistent across all eyes in this analysis. Because the scleral spur is better seen on TD-OCT [6, 7] and could be identified in most of the angle scans, it appeared to represent a better anatomical landmark for assessing ACA. In addition, in this study we enrolled only normal eyes to measure ACA metrics and further studies are warranted to investigate these effects on other groups, especially patients with narrow angles. Finally, we chose only the inferior angle for analysis because we assumed that the dynamic ACA metrics is similar in other quadrants. This assumption has been validated in previous studies [14].

Conclusion

ACA metrics with time domain OCT in normal eyes did not show significant difference from morning to afternoon imaging, or when the angle of incidence was offset by 5° in either direction away from the inferior angle 6 o’clock position. However, there was a significant decrease in angle metrics when the lights were switched off in the imaging room. These finding supports the need to carefully monitor lighting conditions when imaging participants, especially with regards to longitudinal studies.

Abbreviations

ACA:

Anterior chamber angle

AOD:

Angle opening distance

AS:

Anterior Segment

AZ:

Arizona

CA:

California

DIRC:

Doheny Image Reading Center

FC:

Foot candle

IC:

Inferior corneal limbus

OCT:

Optical Coherence Tomography

PACG:

Primary angle closure glaucoma

UBM:

Ultrasound Biomicroscopy

TD:

Time Domain

TISA:

Trabecular-iris space area

References

  1. Quigley HA, Broman AT. The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol. 2006;90(3):262–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Tian J, Marziliano P, Baskaran M, Wong HT, Aung T. Automatic anterior chamber angle assessment for HD-OCT images. IEEE Trans Biomed Eng. 2011;58(11):3242–9.

    Article  PubMed  Google Scholar 

  3. Liu S, Yu M, Ye C, Lam DS, Leung CK. Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. Invest Ophthalmol Vis Sci. 2011;52(12):8598–603.

    Article  PubMed  Google Scholar 

  4. Nolan WP, See JL, Chew PT, Friedman DS, Smith SD, Radhakrishan S, et al. Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology. 2007;114(1):33–9.

    Article  PubMed  Google Scholar 

  5. Leung CK, Li H, Weinreb RN, Liu J, Cheung CY, Lai RY, et al. Anterior chamber angle measurement with anterior segment optical coherence tomography: a comparison between slit lamp OCT and Visante OCT. Invest Ophthalmol Vis Sci. 2008;49(8):3469–74.

    Article  PubMed  Google Scholar 

  6. Dinc UA, Oncel B, Gorgun E, Yalvac IS. Assessment of anterior chamber angle using Visante OCT, slit-lamp OCT and Pentacam. Eur J Ophthalmol. 2010;20(3):531–7.

    PubMed  Google Scholar 

  7. Wylegała E, Teper S, Nowińska AK, Milka M, Dobrowolski D. Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. J Cataract Refract Surg. 2009;35(8):1410–4.

    Article  PubMed  Google Scholar 

  8. Radhakrishnan S, Rollins AM, Roth JE, Yazdanfar S, Westphal V, Bardenstein DS, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001;119(8):1179–85.

    Article  CAS  PubMed  Google Scholar 

  9. Radhakrishnan S, Goldsmith J, Huang D, Yazdanfar S, Westphal V, Bardenstein DS, et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;123:1053–9.

    Article  PubMed  Google Scholar 

  10. Leung C, Weinreb R. Anterior chamber angle imaging with optical coherence tomography. Eye (Lond). 2011;25:261–7.

    Article  Google Scholar 

  11. Akil H, Marion K, Dastiridou A, Jenkins D, Kramer B, Francis BA, et al. Identification of anterior chamber angle parameters with a portable SD-OCT device compared to a non-portable SD-OCT. Int Ophthalmol. 2017;37(1):31–7.

    Article  PubMed  Google Scholar 

  12. Lee JR, Jeoung JW, Choi J, Jin Young C, Park KH, Kim YD. Structure-function relationships in normal and glaucomatous eyes determined by time- and spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(12):6424–30.

    Article  PubMed  Google Scholar 

  13. Xu L, Cao WF, Wang YX, Chen CX, Jonas JB. Anterior chamber depth and chamber angle and their associations with ocular and general parameters: the Beijing Eye study. Am J Ophthalmol. 2008;145(5):929–36.

    Article  PubMed  Google Scholar 

  14. Dastiridou AI, Pan X, Zhang Z, Marion KM, Francis BA, Chopra V. Comparison of physiologic versus pharmacologic mydriasis on anterior chamber angle measurements using spectral domain optical coherence tomography. J Ophthalmol. 2015;2015:845643.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Wang B, Sakata LM, Friedman DS, Chan YH, He M, Lavanya R, et al. Quantitative iris parameters and association with narrow angles. Ophthalmology. 2010;117(1):11–7.

    Article  PubMed  Google Scholar 

  16. Cheung CY, Liu S, Weinreb RN, Liu J, Li H, Leung DY, et al. Dynamic analysis of iris configuration with anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(8):4040–6.

    Article  PubMed  Google Scholar 

  17. Read SA, Collins MJ, Iskander DR. Diurnal variation of axial length, intraocular pressure, and anterior eye biometrics. Invest Ophthalmol Vis Sci. 2008;49(7):2911–8.

    Article  PubMed  Google Scholar 

  18. Clark CV, Mapstone R. Diurnal variation in onset of acute closed angle glaucoma. Br Med J. 1986;292(6528):1106.

    Article  CAS  Google Scholar 

  19. Khor WB, Sakata LM, Friedman DS, Narayanaswamy A, Lavanya R, Perera SA, et al. Evaluation of scanning protocols for imaging the anterior chamber angle with anterior segment-optical coherence tomography. J Glaucoma. 2010;19(6):365–8.

    Article  PubMed  Google Scholar 

  20. Sakata LM, Lavanya R, Friedman DS, Aung HT, Gao H, Kumar RS, et al. Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology. 2008;115(5):769–74.

    Article  PubMed  Google Scholar 

  21. Pan X, Marion KM, Maram J, Zhang ZY, Francis BA, Nittala MG, et al. Reproducibility of anterior segment angle metric measurements derived from cirrus spectral domain optical coherence tomography. J Glaucoma. 2015;24(5):47–51.

    Article  Google Scholar 

  22. Leung CK, Cheung CY, Li H, Dorairaj S, Yiu CK, Wong AL, et al. Dynamic analysis of dark–light changes of the anterior chamber angle with anterior segment OCT. Invest Ophthalmol Vis Sci. 2007;48(9):4116–22.

    Article  PubMed  Google Scholar 

  23. Quigley HA, Silver DM, Friedman DS, He M, Plyler RJ, Eberhart CG, et al. Iris cross-sectional area decreases with pupil dilation and its dynamic behavior is a risk factor in angle closure. J Glaucoma. 2009;18(9):173–9.

    Article  PubMed  Google Scholar 

  24. Aptel F, Chiquet C, Beccat S, Denis P. Biometric evaluation of anterior chamber changes after physiologic pupil dilation using pentacam and anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2012;53(7):4005–10.

    Article  PubMed  Google Scholar 

  25. Stone RA, Quinn GE, Francis EL, Ying GS, Flitcroft DI, Parekh P, et al. Diurnal axial length fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2004;45(1):63–70.

    Article  PubMed  Google Scholar 

  26. Wilson LB, Quinn GE, Ying G, Francis EL, Schmid G, Lam A, et al. The relation of axial length and intraocular pressure fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2006;47(5):1778–84.

    Article  PubMed  Google Scholar 

  27. Pan X, Maram J, Marion K, Dastiridou A, Zhang ZY, Dominguez-Velasco MF, et al. Effect of angle of incidence on anterior chamber angle metrics from optical coherence tomography. J Glaucoma. 2016;25(1):19–23.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Availability of data and materials

Raw data sheet has been submitted along with the manuscript.

Authors’ contributions

HA, AD, KM, BF and VC: Conception and design, acquisition, analysis and interpretation of data, drafting of manuscript, administrative and technical support. HA and KM: Analysis and interpretation of data. VC and BF: Supervision. Read and approved the final manuscript. All authors approved the manuscript for submission.

Competing interests

There are no financial or associative conflicts of interest to disclose by the authors.

Consent for publication

Written informed consent was obtained from all participants to publish this information.

Ethics approval and consent to participate

Written informed consent was obtained from all patients and volunteers before examination. This study was approved by the Institutional Review Board of UCLA Doheny Eye Centers and followed the tenets of the Declaration of Helsinki.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Vikas Chopra.

Additional file

Additional file 1:

Effects of Diurnal, Lighting, and Angle-of-Incidence Variation on Anterior Segment Optical Coherence Tomography Angle Metrics. Description of the data: Angle metrics were recorded from the scans of the morning and the afternoon with different lightining conditions. (XLSX 48 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions