Skip to main content

Table 6 Accuracy of predictability between SRK II and Pediatric IOL Calculator

From: Prediction error and accuracy of intraocular lens power calculation in pediatric patient comparing SRK II and Pediatric IOL Calculator

Observed refraction (Spherical Equivalent)

SRK II (n = 16)

Pediatric IOL Calculator (n = 15)

Overall (n = 31)

* p value

≤ ± 2.0 Diopter

    

   Accurate

    

   0.0 D to ± 0.5 D

3 (18.75%)

7 (46.67%)

10 (32.26%)

0.097

   Inaccurate

    

   > ± 0.5 D to ± 1.0 D

7 (43.75%)

0 (0.00%)

7 (22.58%)

 

   > ± 1.0 D to ± 2.0 D

4 (25.00%)

5 (33.33%)

9 (29.03%)

 

> ± 2.0 Diopter

2 (12.50%)

3 (20.00%)

5 (16.13%)

 
  1. * Pearson chi-square test