Table 1 Biometric Measurements.

From: Comparison of the biometric values obtained by two different A-mode ultrasound devices (Eye Cubed vs. PalmScan): A Transversal, descriptive, and comparative study

Phakic:

Axial Length

Anterior Chamber Deep

Lens Thickness

PalmScan:

23.33 ± 0.97

3.25 ± 0.36

4.52 ± 0.55

Eye Cubed:

23.20 ± 0.97

2.73 ± 0.42

4.59 ± 0.55

p

0.4

<0.0001

0.7

Pseudophakic:

   

PalmScan:

24.52 ± 1.42

4.37 ± 0.47

0.70 ± 0.00

Eye Cubed:

24.44 ± 1.60

3.99 ± 0.47

0.76 ± 0.11

p

0.93

0.1

0.2

  1. Means ± standard deviation in mm. All the patients in the pseudophakic group obtained the same value of lens thickness with the PalmScan™. Due to this, and because there were only six patients in this group, the standard deviation equals zero.