Table 1 The means and standard deviations for ECD measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired t -test, un-paired test, and p-values)

From: Assessment of central corneal thickness and corneal endothelial morphology using ultrasound pachymetry, non-contact specular microscopy, and Confoscan 4 confocal microscopy

Analysis of ECD values

Observer 1

Observer 2

 

SP-2000P

z-ring CS4

z-ring CS4

Mean ± SD*

3115.50 ± 279.70

3167.50 ± 264.75

3192.63 ± 249.42

Intra-observer repeatability (p-value)

0.22

0.99

0.13

 

Mean difference ± SD*

Limits of agreement

p-value

Inter-observer reproducibility with z-ring CS4**

25.10 ± 128.92

+304, -354.20

0.30

Comparison between SP-2000P vs. z-ring CS4†

-52.10 ± 97.90

+139.80, -243.90

0.46

  1. *Standard deviation.
  2. **Paired test.
  3. †Unpaired test.