Table 2 The means and standard deviations for CV measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired t -test, un-paired test, and p-values)

From: Assessment of central corneal thickness and corneal endothelial morphology using ultrasound pachymetry, non-contact specular microscopy, and Confoscan 4 confocal microscopy

Analysis of CV values

Observer 1

Observer 2

 

SP-2000P

z-ring CS4

z-ring CS4

Mean ± SD*

27.12 ± 2.51

27.10 ± 2.41

27.17 ± 2.25

Intra-observer repeatability (p-value)

0.75

0.34

0.98

 

Mean difference ± SD*

Limits of agreement

p-value

Inter-observer reproducibility with z-ring CS4**

-0.20 ± 1.20

+3.80, -4.20

0.30

Comparison between SP-2000P vs. z-ring CS4†

0.20 ± 1.45

+3.00, -2.70

0.79

  1. *Standard deviation.
  2. **Paired test.
  3. †Un-paired test.