Skip to main content

Table 2 In-the-bag stability of single-piece and three-piece IOLs at 3 months

From: Comparisons of the in-the-bag stabilities of single-piece and three-piece intraocular lenses for age-related cataract patients: a randomized controlled trial

 

Group A

(single-piece IOL)

Mean ± SD

Group B

(three-piece IOL)

Mean ± SD

Axial movement

0.07 ± 0.30 mm

−0.22 ± 0.23 mm*

Decentration

221 ± 167 μm

198 ± 165 μm

Tilt

1.06 ± 0.49°

1.01 ± 0.45°

  1. The axial movement, decentration and tilt of the two IOLs are presented in Table 2. At the 3-month postoperative visit, the 3-piece IOLs displayed forward axial movement (−0.22 ± 0.23 mm, p < 0.0001) (*p < 0.05). The single-piece IOLs displayed 221 ± 167 μm decentration, whereas the 3-piece IOLs exhibited 198 ± 165 μm decentration. Notes: IOL intraocular lens, SD standard deviation