| Group A (single-piece IOL) Mean ± SD | Group B (three-piece IOL) Mean ± SD |
p-value |
---|
UCVA | 0.04 ± 0.06 (logMAR) | 0.04 ± 0.08 (logMAR) | 1.00 |
BCVA | 0.02 ± 0.09 (logMAR) | 0.03 ± 0.08 (logMAR) | 1.00 |
SE | 0.15 ± 0.18 D | −0.34 ± 0.15 D | <0.001* |
Total aberration | 0.03 ± 0.06 μm | 0.02 ± 0.07 μm | 0.94 |
- The visual performance of the two IOLs is presented in Table 3. Manifest SE at 3 months postoperatively was 0.15 ± 0.18 D in the single-piece IOL group and −0.34 ± 0.15 D in the three-piece IOL group (p < 0.001)(*p < 0.05) . Notes: IOL intraocular lens, SD standard deviation, UCVA uncorrected visual acuity, BCVA best-corrected visual acuity, SE spherical equivalent