Skip to main content

Table 3 GCIPL thickness (Mean ± SD, μm) determined by two macular cube scanning protocols

From: Reproducibility of macular ganglion cell–inner plexiform layer thickness in normal eyes determined by two different OCT scanning protocols

Parameter

Protocol

1st measurement

2nd measurement

3rd measurement

P a

P b

Average

512 × 128

84.65 ± 5.79

84.50 ± 5.63

84.46 ± 5.65

0.984

0.892

200 × 200

84.61 ± 5.71

84.70 ± 5.69

84.72 ± 5.61

0.994

Minimum

512 × 128

81.80 ± 5.33

81.72 ± 5.13

81.43 ± 5.25

0.927

0.632

200 × 200

81.72 ± 5.87

81.70 ± 5.73

80.91 ± 6.28

0.722

Superotemporal

512 × 128

83.04 ± 5.06

82.76 ± 5.11

82.85 ± 5.21

0.96

0.909

200 × 200

82.67 ± 5.40

82.87 ± 5.37

82.70 ± 5.05

0.977

Superior

512 × 128

85.83 ± 6.10

85.67 ± 6.22

86.00 ± 6.07

0.961

0.974

200 × 200

85.72 ± 6.03

85.80 ± 6.03

85.79 ± 5.98

0.997

Superonasal

512 × 128

87.96 ± 6.56

87.72 ± 6.24

87.83 ± 6.15

0.981

0.963

200 × 200

87.81 ± 6.39

87.83 ± 6.35

87.92 ± 6.27

0.995

Inferonasal

512 × 128

85.22 ± 6.33

85.11 ± 6.24

84.94 ± 6.12

0.973

0.772

200 × 200

85.52 ± 6.22

85.35 ± 6.17

85.70 ± 6.30

0.96

Inferior

512 × 128

82.50 ± 6.32

82.48 ± 6.07

82.20 ± 6.36

0.963

0.726

200 × 200

82.54 ± 6.51

82.61 ± 6.28

83.00 ± 6.52

0.923

Inferotemporal

512 × 128

83.46 ± 5.77

83.39 ± 5.58

83.28 ± 5.72

0.986

0.973

200 × 200

83.35 ± 5.62

83.52 ± 5.69

83.42 ± 5.64

0.988

  1. aComparison between the three repeated measurements of the GCIPL thickness determined by each raster scanning protocol