From: In vitro optical quality measurements of three intraocular lens models having identical platform
Strehl Ratio in 3.0 mm Pupil Size | |||
---|---|---|---|
CT ASPHINA 409 M | AT LISA 809 M | AT LISA Tri 839MP | |
Far | 0.948 | 0.506 | 0.449 |
Intermediate | 0.196 | ||
Near | 0.334 | 0.295 | |
Total SR | 0.948 | 0.840 | 0.940 |
Light Loss | 5.2% | 16.0% | 6.0% |