Skip to main content

Table 3 Summary of correlations between the different anatomical parameters and their within-subject standard deviation (Sw,) and coefficient of variation (CV) associated

From: Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes

 

Sw

CV

 

Pearson correlation coefficient

p-value

Pearson correlation coefficient

p-value

ACD (mm)

0.033

0.734

−0.058

0.550

PCTN3 (μm)

0.024

0.893

−0.082

0.652

PCTN2 (μm)

−0.125

0.487

−0.220

0.218

PCTN1 (μm)

0.101

0.577

0.000

0.999

CCT (μm)

0.149

0.127

0.056

0.565

PCTT1 (μm)

−0.069

0.704

−0.158

0.381

PCTT2 (μm)

−0.059

0.746

−0.142

0.431

PCTT3 (μm)

0.000

0.997

−0.100

0.580

IAN (°)

0.204

0.254

0.122

0.498

IAT (°)

0.039

0.827

−0.102

0.572

WTW (mm)

−0.109

0.282

−0.154

0.125

  1. Abbreviations: SD standard deviation, CV coefficient of variation, S w within-subject standard deviation, Pr intrasubject precision: 1.96 x Sw, ICC intraclass correlation coefficient, ACD anterior chamber depth, PCT N1 , PCT N2 , and PCT N3 peripheral corneal thickness at 1, 2 and 3 mm from vertex nasally, PCT T1 , PCT T2 , and PCT T3 peripheral corneal thickness at 1, 2 and 3 mm from vertex temporally, CCT central corneal thickness, IA N and IA T nasal and temporal irido-corneal angles, and WTW horizontal white-to-white corneal diameter