Skip to main content

Table 1 Summary of findings

From: Goldmann applanation tonometry error relative to true intracameral intraocular pressure in vitro and in vivo

Summary table In vivo (live) +/−SD In vitro (Cadaver) +/−SD
Upright IOP Error (mmHg) -5.2 1.6 −3.1 2.5
Supine IOP Error (mmHg) −7.9 2.3 −5.4 3.1
Difference Upright-Supine (mmHg) −2.7 1.3 −2.3 1.9
CCT Sensitivity (mmHg/μmCCT) 0.024 0.18 0.012 0.1
CRF Sensitivity (mmHg/CRF unit) 0.037 0.08