Table 1 A list of 32 toric intraocular lenses (TIOL) implanted over a 10 months period

From: Real-world refractive outcomes of toric intraocular lens implantation in a United Kingdom National Health Service setting

ID Age (yr) Eye Co-morbidity IOL type IOL power (D) Pre-op BCVA Post-op UVA
1 67 R   Tecnis ZCT 400 20.0 0.30 0.30
2 61 R   Tecnis ZCT 225 14.0 0.30 0.00
L   Tecnis ZCT 225 14.0 0.50 0.10
3a 56 R   Tecnis ZCT 400 12.0 0.30 0.00
4 76 R   Tecnis ZCT 225 18.0 0.50 0.18
5 77 Lb   Tecnis ZCT 150 20.0 0.30 0.00
R   Tecnis ZCT 400 19.0 0.60 0.18
6 67 R   Tecnis ZCT 400 16.5 0.30 0.00
L ERM Tecnis ZCT 400 13.5 0.50 0.48
7a 49 R Vitrectomised Tecnis ZCT 300 19.5 0.50 0.30
8 67 L   Tecnis ZCT 400 16.5 0.30 0.10
R   Tecnis ZCT 225 11.0 0.20 0.00
9 78 R   Tecnis ZCT 300 20.5 0.50 0.30
10 82 L   Tecnis ZCT 225 23.5 0.00 0.00
11 82 R   Tecnis ZCT 400 24.5 0.20 0.18
12 65 L   Tecnis ZCT 400 17.5 0.30 0.00
R   Tecnis ZCT 400 11.0 0.30 0.18
13 41 R   Tecnis ZCT 400 26.5 2.00 0.18
L Amblyopia Tecnis ZCT 400 27.5 1.00 0.18
14 60 L   Tecnis ZCT 400 26.0 0.00 0.30
R   Tecnis ZCT 400 26.5 0.30 0.30
15 80 R Amblyopia Tecnis ZCT 400 27.0 0.60 0.30
16 21 R   Tecnis ZCT 300 22.5 0.48 0.00
17 78 L   Tecnis ZCT 225 31.0 0.18 0.18
18 85 L   Tecnis ZCT 400 21.0 0.30 0.18
19 82 R   Tecnis ZCT 400 17.5 0.30 0.30
20 86 L   Tecnis ZCT 300 23.5 0.30 0.10
21 53 R Previous LASIK & PCRI AcrySof SN60 T8 23.0 0.18 0.10
L AcrySof SN60 T8 22.0 0.30 0.10
22 74 L   Rayner T-flex 623 18.0 0.48 0.18
23 72 L   Rayner T-flex 623 16.0 0.30 0.18
24 74 L Corneal scar Rayner T-flex 623 16.5 0.30 0.18
  1. aIndicate cases complicated by TIOL rotation, which were corrected surgically
  2. bThe left eye of patient 5 received a low powered toric IOL (Tecnis ZCT 150) to correct a ΔK of 1.41 D (which is lower than the ΔK > 2.00 D inclusion criteria). This was at the clinician’s discretion as the patient had already received a higher powered toric IOL in the fellow eye