Table 4 Comparison of examination failure rate between the two devices

From: Comparison of a new swept-source optical biometer with a partial coherence interferometry

 

Number of eyes

Failure rate (%)

Failure

Success

Total

PCI biometer

25

150

175

14.29

SS-OCT biometer

2

173

175

1.14

  1. PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography