Skip to main content

Table 1 Baseline demographics and mean IOP of the study population

From: Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group

 

N

Age

NCT (mm Hg)

iCare Pro (mm Hg)

GAT (mm Hg)

CCT (μm)

AL (mm)

Mean ± SD

range

Mean ± SD

range

Mean ± SD

range

Total

200

56.9 ± 18.3

21.4 ± 12.0

4.0–62.4

21.0 ± 11.5

6.1–62.0

20.7 ± 11.6

4.0–63.3

537.8 ± 36.7

24.1 ± 2.1

Male

76

51.8 ± 19.2

24.9 ± 13.9

4.0–62.4

24.7 ± 14.0

6.5–62.0

24.2 ± 14.2

4.0–63.3

541.2 ± 35.6

24.8 ± 2.5

Female

124

60.1 ± 17.0

19.3 ± 10.2

6.5–56.1

18.7 ± 9.0

6.1–50.8

18.5 ± 9.1

6.0–50.6

535.8 ± 37.4

23.6 ± 1.6

  1. NCT Non-contact tonometer, iCare rebound tonometer iCare Pro, GAT Goldmann tonometry, IOP Intraocular pressure, SD Standard deviation, CCT Central corneal thickness, AL Axial length