Skip to main content

Table 3 Agreement between OLCR and US in measuring AL, ACD and LT

From: Influence of measurement differences of anterior chamber depth and axial length on lens thickness evaluation in cataract patients: a comparison of two tests

Parameter Mean differences SD 95% limits of agreement
AL (mm) 0.27 0.21 −0.15 to 0.70
ACD (mm) 0.15 0.19 −0.22 to 0.52
LT (mm) 0.01 0.44 −0.86 to 0.88