Table 1 Summary of the ocular biometric measurements measured by ANTERION, CASIA2, and IOLMaster 700

From: Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices

 

ANTERION

CASIA2

IOLMaster 700

ANTERION vs IOLMaster 700

CASIA2 vs IOLMaster 700

ANTERION vs CASIA2

ICC

Difference

ICC

Difference

ICC

Difference

Anterior average K, D

43.92 ± 1.62

44.02 ± 1.57

43.98 ± 1.63

0.991

− 0.059 ± 0.213

0.989

0.038 ± 0.233

0.990

− 0.097 ± 0.201

Anterior steep K, D

44.42 ± 1.62

44.51 ± 1.55

44.38 ± 1.60

0.974

0.034 ± 0.367

0.959

0.13 ± 0.439

0.962

−0.096 ± 0.433

Anterior flat K, D

43.42 ± 1.70

43.53 ± 1.66

43.58 ± 1.68

0.976

−0.166 ± 0.334

0.978

−0.059 ± 0.348

0.970

−0.107 ± 0.403

Anterior cylinder K, D

1.01 ± 0.71

0.99 ± 0.76

0.80 ± 0.46

0.598

0.21 ± 0.516

0.443

0.189 ± 0.653

0.526

0.021 ± 0.719

Total average K, D

43.33 ± 1.63

42.89 ± 1.52

43.94 ± 1.61

0.922

−0.61 ± 0.255

0.808

− 1.049 ± 0.237

0.952

0.439 ± 0.232

Total steep K, D

43.85 ± 1.64

43.36 ± 1.57

44.37 ± 1.64

0.920

−0.525 ± 0.425

0.801

− 1.016 ± 0.443

0.918

0.491 ± 0.453

Total flat K, D

42.81 ± 1.69

42.42 ± 1.56

43.51 ± 1.62

0.895

−0.708 ± 0.359

0.790

− 1.094 ± 0.353

0.938

0.386 ± 0.438

Total cylinder K, D

1.04 ± 0.74

0.94 ± 0.73

0.86 ± 0.47

0.505

0.183 ± 0.602

0.441

0.078 ± 0.649

0.461

0.105 ± 0.761

ALa, mm

24.00 ± 1.60

NA

23.99 ± 1.54

0.999

−0.005 ± 0.053

NA

NA

NA

NA

CCT, μm

535.89 ± 34.29

533.60 ± 34.26

535.19 ± 35.04

0.994

0.702 ± 3.735

0.986

−1.596 ± 5.57

0.981

2.298 ± 6.304

ACD, mm

2.62 ± 0.39

2.60 ± 0.48

2.53 ± 0.47

0.990

0.058 ± 0.033

0.982

0.075 ± 0.052

0.994

− 0.015 ± 0.049

LTb, mm

4.63 ± 0.43

4.60 ± 0.55

4.54 ± 0.52

0.915

0.154 ± 0.159

0.963

0.088 ± 0.123

0.961

0.064 ± 0.141

  1. Continuous data are presented as mean ± standard deviation
  2. Differences between two devices were defined by subtracting the latter value from the former value
  3. ACD anterior chamber depth, AL axial length, CCT central corneal thickness, D Diopter, ICC intraclass correlation coefficient, K keratometry, LT lens thickness, NA not available, vs versus
  4. aAL was analyzed with 41 eyes that could be measured by both ANTERION and IOLMaster 700
  5. bLT was analyzed with 44 eyes that could be measured by all three devices