Table 3 Results of Various Studies on Comparison between Mean Axial Length Measured with the IOLMaster and A-scan

From: Axial length of cataract eyes: a comparison of two cohorts over a span of 10 years apart

 

IOL Master 500

(Mean ± SD) (mm)

A-Scan

(Mean ± SD) (mm)

Difference

(mm)

Sample Size

P value

Kiss et al. 2002 [21]

23.7

23.5a

0.2

45

-

Németh et al. 2003 [22]

23.73 ± 2.05

23.3 ± 1.95

0.39 ± 0.36

255

< 0.001

Rose et al. 2003 [23]

23.36 ± 1.24

23.21 ± 1.30

0.15

51

0.0002

Bhatt et al. 2008 [24]

23.97

23.92

0.05

421

-

Roy et al. 2012 [25]

23.43 ± 1.06

23.23 ± 0.98

0.2

31

< 0.0001

Gaballa et al. 2017 [26]

26.18 ± 2.92

26.02 ± 2.99

0.2 ± 0.44

40

0.007

Our study

23.45 ± 0.93

23.38 ± 0.88

0.08 ± 0.13

17

0.023

  1. aImmersion technique used