Skip to main content

Table 1 Level of agreement between image analysis software tools and IOLMaster reference standard: mean differences (with 95% CI around the mean estimate) and lower/upper limits of agreement being ±1.96*SD (all data in mm; right eye)

From: Measuring axial length of the eye from magnetic resonance brain imaging

Comparison to IOLMaster reference (mean 24.1 mm)

Average length

Lower limit of agreement

Mean difference

Upper limit of agreement

   

95% CI lower

Mean

95% CI upper

 

Mango

23.2

−2.6

−1.0

− 0.8

−0.6

0.9

ITK

23.6

−1.8

−0.6

− 0.5

− 0.3

0.9

Carestream

23.9

−1.3

−0.3

−0.1

− 0.0

1.0