From: Digital versus slit-beam marking for toric intraocular lenses in cataract surgery
Digital Group | Manual Group | P value | |
---|---|---|---|
TIA, D (range) | 1.77 ± 0.78 (0.78 to 3.88) | 2.01 ± 0.92 (0.96 to 4.20) | 0.402 |
SIA, D (range) | 1.80 ± 0.88 (0.40 to 3.88) | 1.87 ± 1.03 (0.67 to 4.23) | 0.832 |
DV, D (range) | 0.41 ± 0.44 (0.00 to 1.23) | 0.39 ± 0.39 (0.00 to 1.24) | 0.883 |
AE, degrees | |||
arithmetic mean | 1.89 ± 12.60 | 0.28 ± 5.29 | 0.620 |
(range) | (-22 to 39) | (-9 to 11) | |
absolute mean | 7.11 ± 10.44 | 3.39 ± 3.99 | 0.172 |
(range) | (0 to 39) | (0 to 11) | |
ME, D (range) | 0.03 ± 0.34 (-0.49 to 0.92) | -0.14 ± 0.37 (-1.01 to 0.49) | 0.150 |
CI (range) | 1.01 ± 0.23 (0.5 to 1.40) | 0.91 ± 0.18 (0.6 to 1.21) | 0.164 |
IOS (range) | 0.31 ± 0.36 (0.00 to 1.26) | 0.19 ± 0.18 (0.00 to 0.47) | 0.235 |