Skip to main content

Table 4 The Linear Regression of the IOL Power Difference Calculated by the Barrett Universal II Formula

From: The effect of implantable collamer Lens V4c on ocular biometric measurements and intraocular lens power calculation based on Pentacam-AXL and IOLMaster 500

 

Factors

Unstandardized coefficients

Standard deviation

Standardized coefficients

p

IOL power based on Pentacam-AXL

AL difference

−1.921

0.623

− 0.390

0.004*

ACD difference

0.112

0.280

0.052

0.692

WTW difference

−0.262

0.503

−0.065

0.606

K1 difference

−0.514

0.174

−0.418

0.006*

K2 difference

−0.031

0.092

−0.046

0.739

CCT difference

0.017

0.010

0.233

0.090

IOL power based on IOLMaster 500

AL difference

1.394

1.210

0.150

0.256

ACD difference

−0.362

0.285

−0.171

0.212

WTW difference

0.403

0.339

0.155

0.241

K1 difference

−0.433

0.201

−0.304

0.037*

K2 difference

−0.413

0.153

−0.343

0.010*

  1. *p < 0.05