Skip to main content

Table 5 The Linear Regression of the IOL Power Difference Measured by Pentacam-AXL in 44 Eyes

From: The effect of implantable collamer Lens V4c on ocular biometric measurements and intraocular lens power calculation based on Pentacam-AXL and IOLMaster 500

Factors

Unstandardized coefficients

Standard deviation

Standardized coefficients

p

AL difference

−1.422

1.007

−0.199

0.167

ACD difference

0.142

0.287

0.071

0.624

WTW difference

−0.222

0.512

−0.059

0.667

K1 difference

−0.513

0.176

−0.446

0.006*

K2 difference

−0.039

0.093

−0.063

0.678

CCT difference

0.017

0.010

0.244

0.102