Skip to main content

Table 2 Intra-operative oarameters

From: Comparative evaluation of efficacy and safety of ophthalmic viscosurgical devices in phacoemulsification [ISRCTN34957881]

Parameter

Group 1 n = 19

Group 2 n = 19

Group 3 n = 18

One-way ANOVA, F-Value

P value

Post hoc ANOVA, (Scheffe's test)

Mean phaco energy (%)

0.28 ± 0.06

0.31 ± 0.06

0.38 ± 0.09

6.75

0.0024

I Vs III: p = 0.03

II Vs III: p = 0.05

Mean phaco time (min)

2.0 ± 1.0

1.8 ± 0.8

2.1 ± 0.7

0.026

--

--

Effective phaco time (sec)

33.8 ± 17.7

36.1 ± 20.4

49.5 ± 23.1

3.31

--

--

Mean total surgical time (min)

12.9 ± 3.3

15.2 ± 3.5

16.5 ± 3.6

5.06

 

1 Vs III: p = 0.011

OVD removal time (sec)

66.6 ± 11.2

45.1 ± 9.0

55.47 ± 6.6

26

0.0001

I Vs II: p = 0.0001

I Vs III: p = 0.003

II Vs III: p = 0.006

  1. OVD = Ophthalmic viscosurgical device
  2. min = minutes
  3. sec = seconds