Table 1 Spherical equivalents, sphere, cylinder and axis for each method for the right eye

From: Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia

 

Spherical equivalent

Sphere

Cylinder

Axis

Non-cycloplegic autorefraction (N = 32)

−0.70 (3.14)a

−0.02 (3.33)d

−1.16 (1.35)g

72.75 (66.00)j

 Nidek (N = 21)

0.04 (2.57)

0.76 (2.72)

−1.13 (1.51)

82.19 (72.05)

 Retinomax (N = 11)

−2.10 (3.74)

−1.50 (3.98)

−1.21 (1.05)

54.73 (50.79)

PlusOptix A09

0.54 (1.83)b

1.27 (2.07)e

−1.46 (1.22)h

84.41 (68.56)k

Cycloplegic autorefraction (N = 32)

1.06 (2.04)c

1.77 (2.20)f

−1.41 (0.98)i

73.50 (67.29)l

 Nidek (N = 21)

1.06 (2.29)

1.75 (2.46)

−1.38 (1.09)

75.14 (71.47)

 Retinomax (N = 11)

1.06 (1.57)

1.80 (1.71)

−1.48 (0.79)

70.36 (61.69)

  1. Data are mean (SD)
  2. a P = 0.004 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
  3. b P = 0.02 between autorefractor (Nidek/Retinomax) and PlusOptix A09
  4. c P = 0.04 between PlusOptix A09 and cycloplegic autorefraction
  5. d P < 0.01 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
  6. e P = 0.007 between autorefractor (Nidek/Retinomax) and PlusOptix A09
  7. f P = 0.044 between PlusOptix A09 and cycloplegic autorefraction
  8. g P = 0.09 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
  9. h P = 0.19 between autorefractor (Nidek/Retinomax) and PlusOptix A09
  10. i P = 0.69 between PlusOptix A09 and cycloplegic autorefraction
  11. j P < 0.01 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
  12. k P = 0.16 between autorefractor (Nidek/Retinomax) and PlusOptix A09
  13. l P = 0.28 between PlusOptix A09 and cycloplegic autorefraction
  14. Correlation of refraction value between
  15. -PlusOptix A09 and cycloplegic autorefraction: r = 0.81, p < 0.001
  16. -PlusOptix A09 and non-cycloplegic autorefraction: r = 0.70, p < 0.001
  17. -Non-cycloplegic and cycloplegic autorefraction: r = 0.77, p < 0.001