From: Effects of optical diameter of intraocular lenses with intrascleral fixation on higher-order aberrations

Fig. 5

The tilt aberration and the angle of rotation of the intraocular lenses (IOLs). a The tilt aberrations of the X-60 and X-70 IOLs are not significantly changed within the central 5.2 mm optic diameter but are significantly changed at an extension of 17 mm (n = 5). b The tilt aberrations of the VA-60BBR and VA-70 AD IOLs are not significantly different within the central 5.2 mm optic diameter. c The rotation angle of the X-60 IOL is significantly greater than that of the X-70 IOL only at an extension of 17 mm. d The rotation angle of the VA-60BBR IOL is significantly greater than that of VA-70 AD IOL at extensions from 15 to 17 mm. NS = not significant, (*; P < 0.05, Mann- Whitney U test)