Skip to main content

Table 3 Intra-observer repeatability and inter-observer reproducibility for the PRTa test

From: Validation of the phenol red thread test in a Chinese population

 

Intra-observer repeatability

Inter-observer reproducibility

Experimental group

control group

Experimental group

Control group

n

63

45

63

45

ICCb

0.747

0.723

0.588

0.610

95% CIc for ICC

0.614 to 0.761

0.548 to 0.838

0.401 to 0.728

0.391 to 0.764

P value

 < 0.001

 < 0.001

 < 0.001

 < 0.001

Bland–Altman 95% limits of agreement

-4.008 to 4.468

-5.227 to 4.738

-5.597 to 4.732

-4.908 to 6.041

  1. aPRT phenol re thread, bICC intraclass correlation coefficient, 95% cCI 95% confidence interval for the mean