Skip to main content

Table 2 Repeatability analysis of biometric measurements provided by MS-39

From: Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT

Parameter

ICC (95%CI)

Sw

CoR

CoV (%)

Repeated-measures ANOVA

F

P

Kf (D)

0.9956(0.9945 to 0.9965)

0.079

0.219

0.184

0.82

0.440

Ks (D)

0.9954(0.9942 to 0.9963)

0.091

0.252

0.207

1.36

0.258

Km (D)

0.9970(0.9963 to 0.9976)

0.067

0.186

0.154

0.12

0.889

TCT (µm)

0.9993(0.9992 to 0.9995)

0.880

2.438

0.165

1.66

0.190

CCT (µm)

0.9996(0.9995 to 0.9997)

0.700

1.939

0.131

2.63

0.073

ACD (mm)

0.9988(0.9984 to 0.9990)

0.014

0.039

0.428

2.03

0.132

CV (mm3)

0.9970(0.9962 to 0.9976)

0.183

0.507

0.317

2.34

0.098

CD (mm)

0.9693(0.9620 to 0.9754)

0.066

0.183

0.556

0.50

0.608

  1. ICC = intraclass correlation coefficient; CI = confidence interval; Sw = within subject standard deviation; CoR = coefficient of repeatability; CoV = coefficient of variation; ANOVA = analysis of variance; Kf = central corneal radius of the flat meridian; Ks = central corneal radius of the steep meridian; Km = mean central corneal radius; TCT = thinnest corneal thickness; CCT = central corneal thickness; ACD = anterior chamer depth; CV = corneal volume; CD = corneal diameter