Table 3 Repeatability analysis of biometric measurements provided by MS-39

From: Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT

8Parameter

Bonferroni whirection (P)

 

Bland-Altman [mean difference (95%LoA)]

1st vs. 2nd

1st vs. 3rd

2nd vs. 3rd

 

1st vs. 2nd

1st vs. 3rd

2nd vs. 3rd

Kf (D)

0.5675

1.0000

1.0000

 

0.01(-0.20 to 0.22)

0.00(-022 to 0.22)

-0.01(-0.23 to 0.22)

Ks (D)

0.1891

0.9929

1.0000

 

-0.01(-0.23 to 0.21)

-0.01(-0.29 to 0.27)

0.00(-0.25 to 0.26)

Km (D)

1.0000

1.0000

1.0000

 

0.00(-0.17 to 0.17)

0.00(-0.20 to 0.19)

0.00(-0.19 to 0.19)

TCT (µm)

0.1075

1.0000

1.0000

 

-0.1(-2.3 to 2.0)

-0.1(-2.8 to 2.6)

0.1(-2.4 to 2.6)

CCT (µm)

0.7150

0.0972

0.6944

 

-0.1(-1.7 to 1.6)

-0.1(-2.2 to 1.9)

-0.1(-2.1 to 2.0)

ACD (mm)

0.9098

0.1404

0.9675

 

0.00(-0.02 to 0.02)

0.00(-0.02 to 0.03)

0.00(-0.02 to 0.02)

CV (mm3)

1.0000

0.1510

0.1722

 

-0.01(-0.57 to 0.56)

-0.04(-0.57 to 0.50)

-0.03(-0.43 to 0.38)

CD (mm)

1.0000

1.0000

1.0000

 

0.00(-0.12 to 0.12)

-0.01(-0.23 to 0.22)

0.00(-0.19 to 0.19)

  1. LoA = limits of agreement; Kf = central corneal radius of the flat meridian; Ks = central corneal radius of the steep meridian; Km = mean central corneal radius; TCT = thinnest corneal thickness; CCT = central corneal thickness; ACD = anterior chamer depth; CV = corneal volume; CD = corneal diameter