Epithelial thickness remodeling after small incision lenticule intrastromal keratoplasty in correcting hyperopia measured by RTVue OCT

Abstract

Purpose

To characterize the in vivo corneal epithelial thickness (CET) remodeling profile in a population of eyes after small incision lenticule intrastromal keratoplasty (SMI-LIKE) for hyperopia.

Methods

The CET profile was measured by RTVue-100 Fourier-domain OCT system across the central 6-mm diameter of the cornea of 17 eyes from 12 subjects (five males and seven females) who accepted corneal stromal lens implantation surgery for correcting hyperopia. The CET were measured at positions with a radius of 0–1.0 mm, 1.0–2.5 mm (divided into eight quadrants) and 2.5–3.0 mm (divided into eight quadrants) from the corneal center. Corneal maximum simulated keratometry (Km) was measured by Pentacam HR anterior segment analyzer to analyze CET changes. The examination data of subjects were collected in four time periods, which were preoperative, short-term postoperative (one week after surgery), mid-term postoperative (the last review within 3–6 months after surgery), and long-term postoperative (the last review over 1–2.5 years after surgery). The changes of CET were compared and analyzed in the four time periods.

Results

Mean CET in 0–1.0 mm, 1.0–2.5 mm and 2.5–3.0 mm of the cornea decreased in one week after surgery, respectively, as compared to CET in the preoperative period, which turned from 55.06 ± 0.82 μm、54.42 ± 0.75 μm、53.46 ± 0.60 μm to 51.18 ± 1.05 μm (P = 0.005), 49.38 ± 0.70 μm (P = 0.000), 51.29 ± 0.59 μm (P = 0.025). In the mid-term postoperative period, mean CET in 0–1.0 mm and 1.0–2.5 mm areas kept thinner than mean CET in the preoperative period, CET in 0–1.0 mm is 50.59 ± 0.76 μm (P = 0.000),CET in 1.0–2.5 mm is 50.23 ± 0.57 μm (P = 0.000), while mean CET in 2.5–3.0 mm area recovered to the same thickness as the preoperative level, which is 54.36 ± 0.66 μm (P = 1.000), until the long-term period, CET stabilized in the above doughnut pattern.

Conclusions

After stromal lenticule implantation for hyperopia, CET showed a remodeled form of thinning in the 0–2.5 mm area and thickening in the 2.5–3.0 mm area, and remained stable within one year after surgery.

Peer Review reports

Background

Corneal refractive surgery corrects refractive errors by changing the morphology of cornea, the epithelial thickness remodels after refractive surgery, which can mask the presence of an irregular stromal surface from front surface morphology to a certain extent. The general remodeling trend is that the corneal epithelium will try to compensate and recover to the preoperative corneal morphology, so corneal epithelial remodeling is also related to refractive regression, which has been described by Dr. Dan Z. Reinstein's group [1].

RTVue is an ultra-high speed, high-resolution optical coherence tomography (OCT) (Optovue Inc., Fremont, CA, USA). Compared with in vivo confocal microscopy (IVCM) and very high-frequency digital ultrasound (VHF), OCT demonstrates better patient comfort due to its non-contact method, since examinations which contact cornea could compress the epithelium and lead to thinner results [2].

Stromal keratophakia was developed to correct high hyperopia or aphakia [3]. In the early years, the technique was less established, expensive and resulted in poor visual recovery. With the development of femtosecond lasers, stromal lenticules with precise diopter have been studied and used in treating keratectasia [4, 5], hyperopia [6, 7], presbyopia [8], corneal perforation [9], etc. Currently, an extremely accurate stromal lenticule implantation can be achieved.

Dan Z. Reinstein [10] measured the epithelial thickness profile by Artemis 1 (Ultralink LLC) very highfrequency (VHF) digital ultrasound scanning in a population of normal eyes, and demonstrated that the epithelial thickness is characterized by non-uniformity. On an average, the epithelium was thicker inferiorly than superiorly, and thicker nasally than temporally. As described by Alfred Vogt in 1921 [11], the corneal epithelium has the ability to alter its thickness profile to compensate for the changes in stromal surface curvature in order to try and re-establish a smooth, symmetrical optical surface. However, in refractive surgery, it can lead to refractive regression due to filling the ablation zone. In patients with keratoconus [12] and LASIK [13] correction of hyperopia, the corneal epithelium is reshaped like a "doughnut", and in patients with myopia, the central thickness of the corneal epithelium is thickened.

Hence, we described the corneal epithelium remodeling following hyperopic SMI-LIKE with RTVue OCT, and compared it with the previous conclusion in epithelial remodeling following hyperopic LASIK. We also discussed the effect of corneal epithelium remodeling on refractive regression and evaluated the surgical effect in this study.

Subjects and methods

Patients

In this retrospective study, 17 eyes of 12 patients (five males and seven females) between July 2019 and August 2021 at the Jinan Mingshui Eye Hospital, China, were enrolled. Table 1 shows the demographic data of all subjects. This study was approved by the Ethics Committee of the Jinan Mingshui Eye Hospital Review Board (Ref Ethics/2022/001). The age range of subjects was 11–35 years, with a mean age of 20.67 ± 5.73 years. The inclusion and exclusion criteria were based on Dr. Zhou's research [14], except that we included two teens (11 years old and 16 years old, respectively), who had undergone sequential sessions of perceptual learning for amblyopia [15] in the case of high hyperopia corrected with frame glasses, but had not achieved satisfactory results. They had persistent poor vision with severe visual fatigue, and wished to improve their vision. Each patient (minors with their guardians) provided signed informed consent. The study adhered to the tenets of the Declaration of Helsinki. Spherical equivalent (SE) ranging from 2.25 to 9.50 diopters sphere (DS), with an average of 5.37 ± 1.80DS, median 5.50DS. The cylinder ranged from -0.25 to -3.50 D, with an average of -1.26 ± 0.95D.

Table 1 Demographic data of subjects

Surgical techniques

All surgeries were performed by the same surgeon (YL). Details of the SMI-LIKE procedures has been described in detail in our previous article [16]. The donor eye matching the degree of the recipient was selected before surgery (astigmatism ≤ 0.25D). Small incision lenticule extraction (SMILE) was performed first for the myopic donor. A femtosecond laser (Zeiss, Germany) was used to make the pocket on the recipient’s eye, the diameter of corneal cap was 7.5 mm and its thickness is 120 μm, and the lens diameter is 6.5 mm. The lens was removed and temporarily stored in the equilibrium solution for use. SMILE operation was performed on the recipient hyperopia first, by which, -0.50D and the recipient’s astigmatism were removed according to the parameters. The diameter of the pocket was 7.5 mm, the length of the incision was 4 mm, the corneal cap thickness was 130 μm, diameter of the lens was 6.5 mm, and the lens was removed. The donor corneal lens was soaked with 0.22% riboflavin for 90 s, then washed with normal saline to remove excess riboflavin, and implanted into the corneal pocket of the recipient’s eye. The center of the implanted lens would be carefully aligned with the visual axis.

RTVue OCT and pentacam examination

CET was measured by RTVue-100 Fourier-domain OCT system (Optovue Inc., Fremont, CA,USA) and corneal maximum simulated keratometry (Km) was measured by Pentacam HR anterior segment analyzer (Oculus, Germany). The two instruments operated in a similar manner. All the operations were completed by trained skilled technicians in the dark room. The subject was seated, with jaw and forehead leaned against the headrest. During the measurement, the subjects opened their eyes as wide as possible and stared at the indicator light spot, and the examinations were completed without blinking. Pentacam operations were completed with quality specifications showing “OK”. OCT scan was automatically processed by RTVue-CAM software (Version 6.11; Optovue Inc.), providing the pachymetry (corneal thickness) map and indicated CT and CET ranges at the positions with radii of 0.0–1.0 mm (central zone), 1.0–2.5 mm (divided into eight quadrants) and 2.5–3.0 mm (divided into eight quadrants) from the corneal center, there were totally 17 regions per cornea (Fig. 1).

Fig. 1
figure 1

The pachymetry (corneal thickness) map and the CET ranges at the positions with different radii. T, temporal; N, nasal; S, superior; I, inferior

Visual measurements

All subjects underwent uncorrected distance visual acuity (UDVA) and corrected distance visual acuity (CDVA) at 2.5 m acquired by the standardized logarithmic visual acuity chart. The subjects underwent manifest refraction with both normal and dilated pupil preoperatively and manifest refraction with normal pupil postoperatively.

Statistical analysis

Data analysis was performed using SPSS 26.0 software (International Business Machines, Armonk, NY, USA). Repeated measures analysis of variance was used to compare the CET between four time periods before and after surgery. Comparison between preoperative CET and long-term postoperative CET of different quadrants was performed using paired sample t-test for data conforming to normal distribution, and paired sample rank sum test was used to compare the CET with not normal distribution.

Linear regression analysis was performed between 1) the spherical equivalent refraction of the planted lenses (D) and the central CET with a radius of 0–1.0 mm from the corneal center changes (thinning parts of long-term postoperative CET compared with preoperative CET) (μm); 2) the spherical equivalent refraction of the planted lenses (D) and the maximum simulated keratometry changes between preoperative and long-term postoperative periods (D). The linear regression equation and coefficient of determination (R2) was calculated for each correlation.

The comparison of the maximum simulated keratometry changes between preoperative and long-term postoperative periods in 0 ~ 3D, 3 ~ 6D, 6 ~ 9D subjects was performed using one-way analysis of variance.

Descriptive statistics were performed with Microsoft Excel 2016 (Microsoft Corp, Redmond, Washington). P < 0.05 was considered statistically significant.

Results

Average CET changes in 0–1.0 mm, 1.0–2.5 mm, and 2.5–3.0 mm areas

Mean CET in 0–1.0 mm, 1.0–2.5 mm and 2.5–3.0 mm of the cornea decreased in one week after surgery, respectively, as compared to CET in the preoperative period, which turned from 55.06 ± 0.82 μm、54.42 ± 0.75 μm、53.46 ± 0.60 μm to 51.18 ± 1.05 μm (P = 0.005, R = 93.0%, “R”stands for the rate of change, R = postoperative CET/preoperative CETx100%) 、49.38 ± 0.70 μm (P = 0.000, R = 90.7%) 、51.29 ± 0.59 μm (P = 0.025, R = 96.5%). In the mid-term postoperative period, mean CET in 0–1.0 mm and 1.0–2.5 mm areas kept thinner than mean CET in the preoperative period, CET in 0–1.0 mm is 50.59 ± 0.76 μm (P = 0.000,R = 91.9%),CET in 1.0–2.5 mm is 50.23 ± 0.57 μm (P = 0.000,R = 92.3%), while mean CET in 2.5–3.0 mm area recovered to the same thickness as the preoperative level, which is 54.36 ± 0.66 μm (P = 1.000,R = 101.7%), until the long-term period, CET stabilized in the above doughnut pattern (Table 2).

Table 2 Average CET in 0–1.0 mm, 1.0–2.5 mm, and 2.5–3.0 mm areas (μm)

Figures 2, 3, 4 intuitively sketched the epithelial changes in each area, the postoperative CET, from short-term to long-term, in the location with a radius of 0–1 mm (Fig. 2) and 1.0–2.5 mm (Fig. 3) from the corneal center was decreased compared with the preoperative thickness. CET in the zones with a radius of 2.5–3 mm from the corneal center decreased in short-term postoperative period, except CET in N and IN zones which had almost no change. In the mid-term and long-term periods, CET in all zones with a radius of 2.5–3.0 mm from the corneal center became increasingly thicker, and approached or exceeded the level of the preoperative CET (Fig. 4).

Fig. 2
figure 2

Corneal epithelial thickness distribution in regions with a radius of 0–1.0 mm from the corneal center in different time periods

Fig. 3
figure 3

Corneal epithelial thickness distribution in regions with a radius of 1.0–2.5 mm from the corneal center in different time ranges. T, temporal; N, nasal; S, superior; I, inferior

Fig. 4
figure 4

Corneal epithelial thickness distribution in regions with a radius of 2.5–3.0 mm from the corneal center in different time ranges. T, temporal; N, nasal; S, superior; I, inferior

The CET changes in 17 zones between preoperative and long-term postoperative periods.(17 zones include the 0 ~ 1 mm zone, 8 quadrants in 1 ~ 2.5 mm and 8 quadrants in 2.5 ~ 3.0 mm)

The long-term postoperative central CET with a radius of 0–1.0 mm from the corneal center was 50.35 ± 0.82 µm, which indicated significant decrease compared to the preoperative 55.06 ± 0.82 µm (t = 8.025, P = 0.000). The CET changes in each quadrant before and long-term (≥ 1 year) after the surgery in 1.0 ~ 2.5 mm, 2.5 ~ 3 mm areas is shown in Tables 3, and 4.

Table 3 CET in regions with a radius of 1.0–2.5 mm from the corneal center in preoperative and postoperative hyperopia patients
Table 4 CET in regions with a radius of 2.5–3.0 mm from the corneal center in preoperative and postoperative hyperopia patients

Long-term postoperative CET in all zones with a radius of 1.0–2.5 mm from the corneal center also decreased significantly, as shown in Table 3. In terms of the radius of 2.5–3.0 mm from the corneal center, only CET in S zone showed decrease, but CET in SN, I, IT, T and ST indicated no significant difference compared with preoperative CET. CET in N and IN increased compared to the preoperative level (Table 4).

Correlation between the spherical equivalent refraction of the planted lenses and the central epithelial thickness changes

Figure 5 shows a scatterplot for the spherical equivalent refraction of the planted lenses for the central CET with a radius of 0–1.0 mm from the corneal center changes (thinning parts of long-term postoperative CET compared with preoperative CET) (μm). A statistically significant correlation between the spherical equivalent refraction of the planted lenses and the thinning CET parts in the central zone (R2 = 0.392, P = 0.01) was noted. This indicated that the central corneal epithelial thickness was thinner in eyes after higher refractive implantation.

Fig. 5
figure 5

Correlation between the spherical equivalent refraction of the planted lenses (D) and the central CET with a radius of 0–1.0 mm from the corneal center changes (thinning parts of long-term postoperative CET compared with preoperative CET) (μm). One outlier was eliminated in this chart. The linear regression equations and coefficients of determination (R2) are indicated

Correlation between the spherical equivalent refraction of the planted lenses and the maximum simulated keratometry changes

Figure 6 and Table 5 shows the spherical equivalent refraction of the planted lenses and the maximum simulated keratometry (Km) changes between preoperative and long-term postoperative periods. A statistically significant correlation between the spherical equivalent refraction of the planted lenses and the refraction change of the maximum simulated keratometry (R2 = 0.750, P = 0.000) was noted. This indicated that the cornea after higher refractive implantation has a steeper maximum simulated keratometry.

Fig. 6
figure 6

Correlation between the spherical equivalent refraction(SE) of the planted lenses (D) and the maximum simulated keratometry(Km) changes between preoperative and long-term postoperative periods (D). The linear regression equations and coefficients of determination (R2) are indicated

Table 5 The Km changes between preoperative and long-term postoperative periods corelated with the planted lenses (D)

Visual outcomes after SMI-LIKE

Figure 7A shows the cumulative percentage of UDVA in the last follow-up visit ≥ 1 years postoperatively. A total of 10 eyes (59%) had postoperative UDVA equal to or better than preoperative CDVA. Figure 7B shows the change in the LogMAR lines of CDVA, a total of 11 eyes (65%) had postoperative CDVA equal to or better than preoperative CDVA. Figure 7C shows the scatter plot of the attempted versus achieved SE postoperatively. Patients with low degree tend to be over-corrected while those with high degree tend to be under-corrected, two high degree of hyperopia (+ 7.50D SE and + 9.25D SE) eyes showed a significant undercorrection or refractive regression, the rest is within the forecast. A total of 71% of eyes were within ± 1D in a manifest optometry with normal pupil postoperatively (Fig. 7D).

Fig. 7
figure 7

Visual analyses followed small-incision lenticule intrastromal keratoplasty (SMI-LIKE). A Postoperative uncorrected distance visual acuity (UDVA). B Change in corrected distance visual acuity (CDVA). C Scatter plot of attempt versus achieved spherical equivalent (SE) refraction. D Postoperative SE distribution

Discussion

This is a study to characterize the in vivo CET remodeling over a 6-mm diameter area in eyes after hyperopic SMI-LIKE from one week postoperative to long-term postoperative period more than one year. All eyes showed an epithelial doughnut pattern, in the long-term after surgery, we found an average CET of 50.35 ± 0.82 μm at the 0–1.0 mm area, 49.81 ± 0.69 μm at the 1.0–2.5 mm area and 53.78 ± 0.91 μm at the 2.5–3.0 mm area. Similar to post-LASIK profile, post-SMI-LIKE epithelium also remodeled as a doughnut shape, this is consistent with Zheng Wang, et al.’s conclusion [17]. We wanted to compare it with the post-LASIK profile to explore the long-term effects of different surgical methods for correcting hyperopia, but at our surgical center, patients who chose LASIK to correct hyperopia in recent years were mostly with low to moderate hyperopia accompanied by large astigmatism, so we could not collect relevant data for comparative analysis. Dr. Reinstein’s group studied the epithelial thickness profile by Artemis very high-frequency (VHF) digital ultrasound scanning (ArcScan Inc) over a 10-mm diameter area in 65 eyes of 34 consecutive patients at least three months following hyperopic LASIK with a mean attempted spherical equivalent refraction of + 3.84 ± 1.63 D (range: + 0.75 to + 7.13 D) and a mean attempted cylinder of 1.01 ± 1.04 D (range: 0.00 to 5.00 D), and found the average CET was 39.7 ± 5.6 μm at the central thinnest location and 89.3 ± 14.6 μm at the paracentral thickest location, while the mean postoperative maximum simulated keratometry was 48.10 ± 2.10 D (range: 43.20 to 52.60 D) [13]. In this study, in the long-term postoperative period, the thinnest CET in the central area was 48.18 ± 2.86 μm located in S quadrant of the 1.0–2.5 mm annulus, and the thickest CET was 58.06 ± 5.56 μm located in the IN quadrant of the 2.5–3.0 mm annulus, which showed a flatter doughnut pattern of the epithelium in treatment of larger degrees of hyperopia, regardless of differences in race, sample size and measurement methods. Meanwhile, this uneven CET distribution is consistent with CET distribution in the normal cornea, which may be related to the friction of the eyelid against the epithelium during blinking [10].

In the patient with a + 9.50D lenticule implantation, the maximum thinning of epithelial thickness was located in zone I of the 1–2.5 mm ring, which was 13 μm thinner than before surgery. This change was much smaller than that reported by Dr. Reinstein after LASIK hyperopia, and similar to the epithelial thickening after myopia ablation [13, 18]. This could be explained by the epithelial response according to the corneal morphological changes; hyperopic ablations by LASIK form a steep shape in cornea passively by ablating the peripheral optical zone, the corneal shaping after SMI-LIKE is more gradual and on its own. The epithelium appears to compensate more for a localized stromal irregularity.

Corneal topography is useful in predicting the final keratometry after refractive surgery. Central keratometry steeper than 49 to 50 D after LASIK may be associated with a decrease in quality of vision [19]. In our study, the mean postoperative maximum simulated keratometry was 46.68 ± 1.67 D (range: 43.70 to 49.2 D).

Irrespective of the surgical method, the main concerns are the postoperative effect and when to achieve stability. LASIK following hyperopia usually achieved stability within 1–2 years after surgery [20,21,22]. It has a good therapeutic effect for low hyperopia, but more frequent worse predictability and a loss of best spectacle-corrected visual acuity (BSCVA) [19, 23, 24] in treating higher degrees of hyperopia. Moreover, surgical pattern and optical zone also play roles in the surgical effect [25, 26].

In the visual outcomes, some patients showed over-corrected, we considered that patients with hyperopia are usually accompanied by strong hyper-regulation, most of them have already adapted to about -2.0D of accommodative myopia, so they presented over-corrected postoperatively. Some patients were reserved some degree of hyperopia in the design of preoperatively surgical parameters, because patients with >  + 8.0D hyperopia were difficult to find enough matching donors, and the two minors were retained some degree considering the increase of the eye axial length during growth, so in Fig. 7D, 18% of total showed > 2.00D SE(+ 2.00D ~  + 3.00D) in a ≥ 1 year follow-up visit. The 11-year-old patient’s eyesight turned from a UDVA1.0, CDVA0.1 preoperatively to UDVA 0.4, CDVA0.2 one year after surgery, and the 16-year-old patient went from UDVA1.2, CDVA1.2 preoperatively to UDVA0.9, CDVA0.9 one year after surgery. Patients might get a better vision with a sequential sessions of perceptual learning for amblyopia postoperatively, but most of them refused that due to lack of time or uncertain efficacy.

According to the current results, there is no significant statistical difference in corneal epithelial thickness between the mid-term and the long-term postoperative periods, indicating that the corneal epithelium can reach a stable state within half a year after surgery. Epithelial remodeling and corneal biomechanics both play significant roles in hyperopic refractive regression [27], and the CET changes following SMI-LIKE suggest that this type of surgery may have lower refractive regression rate and level. Refractive regression is not only related to corneal epithelial remodeling, but also to stroma changes after surgery. Following previous studies, some changes have become predictable, thereby improving the accuracy of surgery. In SMILE surgeries, a 10% refractive sphere overcorrection nomogram is commonly used to obtain accurate outcomes [28]. Alió del Barrio et al. considered that some central corneal stroma expanded due to the biomechanical changes induced by SMILE [29], as the division of the residual bed and the cap might cause expansion of the stroma because they are no longer under tension [30, 31]. Based on this theory, there may have a overcorrection after SMI-LIKE surgery due to the expansion of the corneal stroma, but whether the expansion is partially offset by the compression of the implanted stromal lens worth further investigation.

We plan to collect the data of diopter, curvature, visual acuity and stroma changes in different postoperative periods in the future to discuss the safety and stability of SMI-LIKE correction of hyperopia.

In conclusion, after stromal lenticule implantation for hyperopia, CET showed a remodeled form of thinning in the 0–2.5 mm area and thickening in the 2.5–3.0 mm area, and remsained stable within one year after surgery. The higher the degree, the more prone to gothrough refractive regression or undercorrection, prolonging postoperative corticosteroid drops use may be useful in patients with high hypermetropia to prevent refractive regression.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

References

  1. Reinstein DZ, Archer TJ, Vida RS. Epithelial thickness mapping for corneal refractive surgery. Curr Opin Ophthalmol. 2022;33(4):258–68. https://doi.org/10.1097/ICU.0000000000000867.

    Article  PubMed  Google Scholar 

  2. Hong JX, Qian TT, Yang YJ, et al. Corneal epithelial thickness map in long-term soft contact lenses wearers. Optom Vis Sci. 2014;91(12):1455–61. https://doi.org/10.1097/OPX.0000000000000410.

    Article  Google Scholar 

  3. Riau AK, Liu YC, Yam GHF, Mehta JS. Stromal keratophakia: Corneal inlay implantation. Prog Retin Eye Res. 2020;75:100780. https://doi.org/10.1016/j.preteyeres.2019.100780.

    Article  PubMed  Google Scholar 

  4. Sachdev MS, Gupta D, Sachdev G, et al. Tailored stromal expansion with a refractive lenticule for crosslinking the ultrathin cornea. J Cataract Refract Surg. 2015;41:918–23.

    Article  Google Scholar 

  5. Cagini C, Riccitelli F, Messina M, et al. Epi-off-lenticule-on corneal collagen cross-linking in thin keratoconic corneas. Int Ophthalmol. 2020;40:3403–12.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Sun L, Yao P, Li M, et al. The safety and predictability of implanting autologous lenticule obtained by SMILE for hyperopia. J Refract Surg. 2015;31:374–9.

    Article  PubMed  Google Scholar 

  7. Pradhan KR, Reinstein DZ, Carp GI, et al. Femtosecond laser-assisted keyhole endokeratophakia: correction of hyperopia by implantation of an allogeneic lenticule obtained by SMILE from a myopic donor. J Refract Surg. 2013;29:777–82.

    Article  PubMed  Google Scholar 

  8. Jacob S, Kumar DA, Agarwal A, et al. Preliminary evidence of successful near vision enhancement with a new technique: PrEsbyopic Allogenic Refractive Lenticule (PEARL) corneal inlay using a SMILE lenticule. J Refract Surg. 2017;33:224–9.

    Article  PubMed  Google Scholar 

  9. Wu F, Jin X, Xu Y, et al. Treatment of corneal perforation with lenticules from small incision lenticule extraction surgery. Cornea. 2015;34:658–63.

    Article  PubMed  Google Scholar 

  10. Reinstein DZ, Archer TJ, Gobbe M, Silverman RH, Coleman DJ. Epithelial thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound. J Refract Surg. 2008;24(6):571–81. https://doi.org/10.3928/1081597X-20080601-05.

    Article  PubMed Central  Google Scholar 

  11. Vogt A. Textbook and atlas of slit lamp microscopy of the living eye. Bonn: Wayenborgh Editions; 1981.

    Google Scholar 

  12. Reinstein DZ, Archer TJ, Gobbe M. Corneal epithelial thickness profile in the diagnosis of keratoconus. J Refract Surg. 2009;25(7):604–10. https://doi.org/10.3928/1081597X-20090610-06.

    Article  Google Scholar 

  13. Reinstein DZ, Archer TJ, Gobbe M, Silverman RH, Coleman DJ. Epithelial thickness after hyperopic LASIK: three-dimensional display with Artemis very high-frequency digital ultrasound. J Refract Surg. 2010;26(8):555–64. https://doi.org/10.3928/1081597X-20091105-02.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Liu S, Zhang X, Zhou X. Toric lenticule implantation for correction of hyperopia and astigmatism following small incision lenticule intrastromal keratoplasty with the triple marking method. J Refract Surg. 2022;38(2):82–8. https://doi.org/10.3928/1081597X-20211117-01.

    Article  PubMed  Google Scholar 

  15. Polat U, Ma-Naim T, Spierer A. Treatment of children with amblyopia by perceptual learning. Vision Res. 2009;49(21):2599–603. https://doi.org/10.1016/j.visres.2009.07.008.

    Article  PubMed  Google Scholar 

  16. Hou J, Wang Y, Zhang J, Lei Y, Ma Z, Zhang Y, Zheng X. Corneal densitometry after allogeneic small-incision intrastromal lenticule implantation for hyperopia correction. BMC Ophthalmol. 2022;22(1):286. https://doi.org/10.1186/s12886-022-02454-3.

    Article  PubMed Central  Google Scholar 

  17. Wu J, Xiong L, Wang Z, Reinstein DZ, Vida RS, Archer TJ. Correction of moderate to high hyperopia with implantation of an allogeneic refractive lenticule. J Refract Surg. 2020;36(11):772–9. https://doi.org/10.3928/1081597X-20200826-01.

    Article  Google Scholar 

  18. Reinstein DZ, Archer TJ, Gobbe M. Change in epithelial thickness profile 24 hours and longitudinally for 1 year after myopic LASIK: three-dimensional display with Artemis very high-frequency digital ultrasound. J Refract Surg. 2012;28(3):195–201. https://doi.org/10.3928/1081597X-20120127-02.

    Article  PubMed  Google Scholar 

  19. Varley GA, Huang D, Rapuano CJ, et al. LASIK for hyperopia, hyperopic astigmatism, and mixed astigmatism: a report by the American Academy of Ophthalmology. Ophthalmology. 2004;111(8):1604–17. https://doi.org/10.1016/j.ophtha.2004.05.016.

    Article  PubMed  Google Scholar 

  20. Kanellopoulos AJ. Topography-guided hyperopic and hyperopic astigmatism femtosecond laserassisted LASIK: long-term experience with the 400 Hz eye-Q excimer platform. Clin Ophthalmol. 2012;6:895–901.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Plaza-Puche AB, Yebana P, Arba Mosquera S, Alió JL. Three-year follow-up of hyperopic LASIK using a 500-Hz excimer laser system. J Refract Surg. 2015;31:674–82.

    Article  PubMed  Google Scholar 

  22. Alió J, Galal A, Ayala MJ, Artola A. Hyperopic LASIK with Esiris/Schwind technology. J Refract Surg. 2006;22:772–81.

    Article  PubMed  Google Scholar 

  23. Choi RY, Wilson SE. Hyperopic laser in situ keratomileusis: primary and secondary treatments are safe and effective. Cornea. 2001;20(4):388–93. https://doi.org/10.1097/00003226-200105000-00010.

    Article  CAS  PubMed  Google Scholar 

  24. Cobo-Soriano R, Llovet F, González-López F, Domingo B, Gómez-Sanz F, Baviera J. Factors that influence outcomes of hyperopic laser in situ keratomileusis. J Cataract Refract Surg. 2002;28(9):1530–8. https://doi.org/10.1016/s0886-3350(02)01367-6.

    Article  PubMed  Google Scholar 

  25. Tham VM, Maloney RK. Microkeratome complications of laser in situ keratomileusis. Ophthalmology. 2000;107(5):920–4. https://doi.org/10.1016/s0161-6420(00)00004-x.

    Article  CAS  PubMed  Google Scholar 

  26. Esquenazi S, Mendoza A. Two-year follow-up of laser in situ keratomileusis for hyperopia. J Refract Surg. 1999;15(6):648–52. https://doi.org/10.3928/1081-597X-19991101-08.

    Article  CAS  PubMed  Google Scholar 

  27. Yan MK, Chang JS, Chan TC. Refractive regression after laser in situ keratomileusis. Clin Exp Ophthalmol. 2018;46(8):934–44. https://doi.org/10.1111/ceo.13315.

    Article  PubMed  Google Scholar 

  28. Wu F, Yin H, Chen X, Yang Y. Investigation of predictability and influence factors of the achieved lenticule thickness in small incision lenticule extraction. BMC Ophthalmol. 2020;20(1):110.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Alió Del Barrio JL, Canto-Cerdan M, El Bahrawy M, Casanova L, García MJ, Cavas F, Alió JL. Corneal stromal thickness changes after myopic laser corneal refractive surgery. J Cataract Refract Surg. 2022;48(3):334–41. https://doi.org/10.1097/j.jcrs.0000000000000765.

    Article  PubMed  Google Scholar 

  30. Reinstein DZ, Archer TJ, Gobbe M. Lenticule thickness readout for small incision lenticule extraction compared to artemis three-dimensional very high-frequency digital ultrasound stromal measurements. J Refract Surg. 2014;30(5):304–9.

    Article  Google Scholar 

  31. Roberts C. The cornea is not a piece of plastic. J Refract Surg. 2000;16(4):407–13.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

Supported by Science and Technology Planning Project of Jinan Health Commission (Grant No. 2022–2-191).

Author information

Authors and Affiliations

Authors

Contributions

Yulin Lei contributed to the study concept and design. Jing Zhang, Xinghua Yang, Fangfang Sun, Yahui Dong acquired and interpreted study data. Yahui Dong conducted statistical analysis and drafted the manuscript.  Yulin Lei and Jie Hou provided critical manuscript revisions and administrative, technical support. Yulin Lei supervised the study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yulin Lei.

Ethics declarations

Ethics approval and consent to participate

The study protocol conformed to the tenets of Declaration of Helsinki, and was approved by the Human Ethics Committee of Jinan Mingshui Eye Hospital (Ref Ethics/2022/001). Informed consent was obtained from all of the adult study participants, the minor participants and their guardians.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dong, Y., Hou, J., Zhang, J. et al. Epithelial thickness remodeling after small incision lenticule intrastromal keratoplasty in correcting hyperopia measured by RTVue OCT. BMC Ophthalmol 24, 13 (2024). https://doi.org/10.1186/s12886-023-03272-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-023-03272-x

Keywords