Table 3 The means and standard deviations for CCT measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired t -test, un-paired test, and p-values)

From: Assessment of central corneal thickness and corneal endothelial morphology using ultrasound pachymetry, non-contact specular microscopy, and Confoscan 4 confocal microscopy

Analysis of CCT values

Observer 1

Observer 2

 

USP

SP-2000P

z-ring CS4

z-ring CS4

Mean ± SD*

555.11 ± 35.83

535.82 ± 41.10

552.57 ± 36.83

554.97 ± 36.34

Intra-observer repeatability (p-value)

0.91

0.20

0.45

0.54

 

Mean difference ± SD*

Limits of agreement

p-value

Inter-observer reproducibility with z-ring CS4**

-2.40 ± 9.10

+39.90, -38.70

0.16

USP vs. SP-2000P

19.30 ± 19.60

+57.70, -19.20

 

USP vs. z-ring CS4

2.50 ± 5.70

+16.30, -11.20

SP-2000P vs. z-ring CS4

-16.70 ± 18.60

+23.70, -57.20

 
  1. Standard deviation*.
  2. Paired test**.