Skip to main content

Table 2 Changes in IOP measurements (n = 22)

From: Changes in intraocular pressure values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology tonometer (CST) in the early phase after small incision lenticule extraction (SMILE)

Variables

Pre-operative

Postoperative 20 min

Postoperative 24 h

F valuea

P value

Mean ± SD

Range

Mean ± SD

Range

Mean ± SD

Range

NCT-IOP (mmHg)

15.04 ± 2.91

11.1 to 20.9

9.19 ± 2.52

4.6 to 15.2

9.70 ± 2.95

5.8 to 17.2

81.173

<0.001b

IOPcc (mmHg)

14.99 ± 2.46

10.2 to 19.0

16.65 ± 4.29

12.1 to 29.4

14.01 ± 3.10

6.8 to 20.3

4.975

0.030b

IOPg (mmHg)

15.10 ± 3.21

9.3 to 21.9

11.20 ± 3.38

4.7 to 18.2

9.52 ± 2.85

4.5 to 15.9

40.128

<0.001b

CST-IOP (mmHg)

17.74 ± 2.16

13.0 to 23.0

14.29 ± 2.09

10.0 to 17.5

14.57 ± 1.58

10.5 to 18.0

31.886

<0.001b

  1. IOPcc corneal-compensated IOP, IOPg Goldmann-correlated IOP
  2. aRepeated measures analysis of variance (RM-ANOVA)
  3. bSignificant difference was detected over time