Skip to main content

Table 6 Comparison of Strehl Ratio and Light Loss values measured at 4.5 mm pupil size

From: In vitro optical quality measurements of three intraocular lens models having identical platform

 

Strehl Ratio in 4.5 mm Pupil Size

 

CT ASPHINA 409 M

AT LISA 809 M

AT LISA Tri 839MP

Far

0.845

0.346

0.240

Intermediate

  

0.111

Near

 

0.279

0.192

Total SR

0.845

0.625

0.543

Light Loss

15.5%

37.5%

45.7%