Table 3 Comparison of V4c ICL size, sulcus-to-sulcus diameter, and postoperative vault between the “normal-sizing” and “under-sizing” groups

From: Buffering zone of implantable Collamer lens sizing in V4c

 

Normal-sizing (N = 19 eyes)

Under-sizing (N = 24 eyes)

Overall (N = 43 eyes)

PNormal_Under

Pre-converted ICL size (mm)

12.11 ± 0.36

12.13 ± 0.40

12.12 ± 0.38

0.866

Pre-converted ICL size – STS (μm)

255.79 ± 248.20

−182.33 ± 113.61

11.26 ± 286.25

0.000

Actual ICL size – STS (μm)

892.63 ± 262.36

463.50 ± 112.75

653.12 ± 288.03

0.000

Actual ICL size (mm)

12.74 ± 0.40

12.77 ± 0.44

12.76 ± 0.42

0.827

Postop vault (μm)

596.32 ± 308.64

535.42 ± 204.51

562.33 ± 254.29

0.442

  1. Actual ICL size: V4c ICL size in balanced salt solution (BSS)
  2. Pre-converted ICL size: ICL size in normal saline before conversion to actual V4c ICL size
  3. PNormal_Under: P-values are given between the “normal-sizing” and “under-sizing” groups
  4. Abbreviations: STS sulcus-to-sulcus diameter, Postop postoperative