Skip to main content

Table 2 Comparison of toric IOL implantation studies

From: Real-world refractive outcomes of toric intraocular lens implantation in a United Kingdom National Health Service setting

Study

Eyes

Age (years ±SD or range)

Toric IOL

Post-op UVA (logMAR) ± SD

Pre-op cyl (D) ± SD

Post-op cyl (D) ± SD

Max IOL rotation (°)

This study

32

66.6 ± 14.5

Tecnis Toric (n = 27), Rayner T-flex (n = 2), Alcon Acrysof toric (n = 3)

0.16 ± 0.12

2.45 ± 1.2

1.04 ± 0.79

30

Tsinpoulos et al. [10]

29

63 ± 5.4

Alcon Acrysof toric

≥0.30 (90%)

2.38 ± 0.91

0.64 ± 0.61

8.4

≥0.10 (66%)

Hirnschall et al [9]

30

67 (36–85)

Tecnis Toric

−0.05

1.80 ± 0.50

0.90 ± 0.40

13.7

Entabi et al. [11]

33

81 ± 8.9

Rayner T-flex

0.28 ± 0.23

3.35 ± 1.20

0.95 ± 0.66

17