Skip to main content

Table 3 Correlations between IOP, as obtained from tested tonometers, and age, spherical equivalent (SE), corneal curvature (MK) and central corneal thickness (CCT)

From: Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices

Correlations

  

AGE

SE

MK

CCT

DCT

Spearman’s rho

−0.117

0.025

−0.067

0.470

Sig. (2-tailed)

0.159

0.762

0.422

0.0001

N

146

146

146

146

GAT

Spearman’s rho

−0.143

0.102

− 0.043

0.625

Sig. (2-tailed)

0.085

0.221

0.608

0.0001

N

146

146

146

146

ORA

Spearman’s rho

−0.089

−0.029

− 0.096

0.413

Sig. (2-tailed)

0.283

0.732

0.251

0.0001

N

146

146

146

146

RT

Spearman’s rho

−0.114

0.072

−0.098

0.519

Sig. (2-tailed)

0.172

0.389

0.240

0.0001

N

146

146

146

146

CST

Spearman’s rho

−0.104

0.040

−0.201

0.522

Sig. (2-tailed)

0.210

0.633

0.015

0.0001

N

146

146

146

146

  1. Legend: Corvis ST (CST), Ocular Response Analyzer (ORA), rebound tonometry (RT), dynamic contour tonometer (DCT) and Goldmann applanation tonometer (GAT). Highlighted values are the correlations resulted to be significant