Table 1 Repeatability of Pentacam, UBM and AS-OCT in measuring ACD, C-ICL and Vault

From: Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation

Device

Parameter

Pearson (r)

P-value

Difference(mm)

t-value

P-value

AS-OCT

ACD

0.98

0.000

0 ± 0.04

−1.20

0.23

C-ICL

0.99

0.000

0 ± 0.04

−0.48

0.64

Vault

0.99

0.000

0 ± 0.03

1.44

0.15

Pentacam

ACD

0.98

0.000

0 ± 0.04

−0.85

0.40

C-ICL

0.99

0.000

0. ± 0.04

−0.90

0.37

Vault

0.99

0.000

0 ± 0.03

−0.77

0.45

UBM

ACD

0.98

0.000

0 ± 0.05

0.36

0.72

C-ICL

0.99

0.000

0 ± 0.05

0.83

0.41

Vault

0.99

0.000

0 ± 0.04

0.22

0.83

  1. AS-OCT: Anterior segment optical coherence tomography; UBM: Ultrasound biomicroscopy; Pearson (r): Pearson correlation coefficient; ACD: Anterior chamber depth; C-ICL: Distance between the corneal endothelium and anterior surface of ICL; Vault: Distance between the back surface of ICL and the front surface of the lens. P values less than 0.05 were considered statistically significant