Table 3 Results of Square Localization (SL) and Direction of Motion (DOM) Testing at month 3 (M3), month 6 (M6), and year 1 (Y1) measured as a change in pixels (SL) or degrees (DOM) between testing with the device OFF and ON (ΔOFF-ON). A positive number indicates an improvement with the device ON

From: Effect of the electrode array-retina gap distance on visual function in patients with the Argus II retinal prosthesis

 

Square Localization Testing

Direction of Motion Testing

ΔOFF-ON (pixels)

ΔOFF-ON (degrees)

Patient #

M3

M6

Y1

Average

M3

M6

Y1

Average

1

18.67

81.47

71.14

57.09

2.39

14.74

14.20

10.45

2

0.90

− 37.71

− 81.36

−39.39

− 15.58

−13.87

−43.74

−24.39

3

− 52.20

−71.52

−52.72

−58.81

−10.03

−19.02

−8.94

−12.66

4

111.50

113.72

100.17

108.46

−2.10

0.63

−21.80

−7.76

5

27.96

18.35

108.99

51.76

1.32

8.54

−3.21

2.22