Skip to main content

Table 8 Test-retest reliability of optic disc and macular scans of glaucomatous and healthy eyes using AngioVue

From: Comparison of two different optical coherence tomography angiography devices in detecting healthy versus glaucomatous eyes – an observational cross-sectional study

Eyes

Scan

Layers

Overall

mean

Sw

Sw 95% CI

CV

CRw

CRw 95% CI

Glaucoma

Optic disc

Vitreous

0.054

0.012

0.009–0.015

22

0.033

0.025–0.041

Radial peripapillary capillaries

0.050

0.016

0.013–0.020

33

0.046

0.035–0.057

Nerve head

0.095

0.016

0.013–0.020

17

0.045

0.035–0.055

Choroid

0.178

0.027

0.021–0.033

15

0.075

0.057–0.092

Macula

Superficial retina

0.074

0.007

0.005–0.008

9

0.018

0.014–0.023

Deep retina

0.106

0.011

0.008–0.013

10

0.030

0.022–0.037

Outer retina

0.083

0.012

0.009–0.015

14

0.033

0.025–0.041

Choroid

0.128

0.010

0.007–0.012

7

0.027

0.020–0.033

Control

Optic disc

Vitreous

0.103

0.027

0.022–0.031

26

0.074

0.062–0.087

 

Radial peripapillary capillaries

0.135

0.022

0.019–0.026

17

0.062

0.052–0.072

Nerve head

0.197

0.029

0.025–0.034

15

0.081

0.068–0.094

Choroid

0.264

0.042

0.035–0.049

16

0.117

0.098–0.137

Macula

Superficial retina

0.118

0.016

0.013–0.019

14

0.045

0.036–0.053

Deep retina

0.251

0.017

0.013–0.020

7

0.046

0.037–0.055

Outer retina

0.188

0.027

0.022–0.032

14

0.075

0.061–0.089

 

Choroid

0.255

0.020

0.016–0.024

8

0.056

0.045–0.066

  1. Sw Within-subject standard deviation; CI Confidence interval; CV Coefficient of variation; CRw Coefficient of repeatability