Skip to main content

Table 9 Test-retest reliability of optic disc and macular scans of glaucomatous and healthy eyes using SweptSource

From: Comparison of two different optical coherence tomography angiography devices in detecting healthy versus glaucomatous eyes – an observational cross-sectional study

Eyes

Scan

Layers

Overall

mean

Sw

Sw 95% CI

CV

CRw

CRw 95% CI

Glaucoma

Optic disc

Superficial retina

0.277

0.121

0.093–0.149

44

0.335

0.258–0.412

Deep retina

0.278

0.117

0.090–0.143

42

0.323

0.248–0.398

Choroid

0.432

0.125

0.097–0.153

29

0.347

0.268–0.425

Macula

Superficial retina

0.245

0.120

0.092–0.148

49

0.332

0.254–0.410

Deep retina

0.254

0.117

0.090–0.145

46

0.325

0.249–0.401

Choroid

0.389

0.114

0.088–0.140

29

0.316

0.243–0.389

Control

Optic disc

Superficial retina

0.234

0.065

0.055–0.075

28

0.181

0.153–0.209

Deep retina

0.241

0.065

0.055–0.075

27

0.181

0.153–0.209

Choroid

0.347

0.068

0.058–0.079

20

0.189

0.160–0.219

Macula

Superficial retina

0.226

0.111

0.094–0.128

49

0.307

0.259–0.355

Deep retina

0.241

0.115

0.097–0.133

48

0.320

0.270–0.369

Choroid

0.333

0.106

0.089–0.122

32

0.294

0.248–0.339

  1. Sw Within-subject standard deviation; CI Confidence interval; CV Coefficient of variation; CRw Coefficient of repeatability