Skip to main content

Table 2 Overview of the glistening densities and the Miyata grading [14] of all six intraocular lenses

From: In-vitro glistening formation in six different foldable hydrophobic intraocular lenses

IOL Model

Glistenings/mm2 (mean ± standard deviation)

Miyata grade (mean)

Hoya Vivinex™ XY1

11.6 ± 5.7

0

AMO Tecnis® ZCB00

6.0 ± 2.8

0

Kowa Avansee™ PN6A

2.2 ± 0.7

0

Alcon AcrySof® SN60WF

264.4 ± 110.3

2.6

Zeiss CT LUCIA® 601P

71.0 ± 71.6

1

Nidek Aktis SP NS-60YG

851.4 ± 59.4

3+

  1. IOL Intraocular lens