Skip to main content

Table 3 Intra-rater repeatability of the MRI measures of axial lengths (mm) in right eye measured by two analysts at two time points: agreement by intraclass correlation coefficient

From: Measuring axial length of the eye from magnetic resonance brain imaging

Software

Analyst

1st time point

2nd time point

Difference

ICC (95% CI)

Mango

1

23.2 ± 1.29

23.3 ± 1.31

− 0.1

0.93 (0.89 to 0.95)

2

22.8 ± 1.41

22.8 ± 1.51

0.0

0.90 (0.86 to 0.94)

ITK

1

23.6 ± 1.17

23.5 ± 1.13

0.1

0.95 (0.92 to 0.96)

2

23.5 ± 1.22

23.7 ± 1.21

−0.2

0.90 (0.82 to 0.94)

Carestream

1

23.9 ± 1.13

24.1 ± 1.13

−0.2

0.95 (0.86 to 0.97)

2

23.9 ± 1.13

23.8 ± 1.11

0.1

0.95 (0.93 to 0.97)