Skip to main content

Table 1 Achieved depth of focus (DoF) range and increase (ΔDoF) in Diopters of an enhanced monofocal in comparison to a standard monofocal intraocular lens for predicted visual acuity (sVA) at 0.2 logMAR estimated from optical quality metrics

From: Positioning of enhanced monofocal intraocular lenses between conventional monofocal and extended depth of focus lenses: a scoping review

Author (Year)

Measurement device

Spherical aberration of cornea

Enhanced Monofocal

Predictor Metric

Aperture (mm)

DoF Range [ΔDoF]

Alarcon et al. (2020) [18]

Optical Bench

Unspecified + Z40

Tecnis Eyhance

sVA (0.2)

3

1.73 [0.5]

Vega et al. (2020) [19]

Optical Bench

 + 0.18 µm at 5 mm

Tecnis Eyhance

MTFa (11.4)a

3

1.89 [0.3]

Labuz et al. (2021) [20]

OptiSpheric

 + 0.28 µm at 5.15 mm

RayOne EMV

wOTF (0.51)b

3

1.49 [0.07]

Azor et al. (2022) [15]

Optical Bench

 + 0.18 µm at 5 mm

Tecnis Eyhance

MTFa (11.4)a

3

1.83 [0.3]

Azor et al. (2022) [15]

Optical Bench

 + 0.18 µm at 5 mm

IsoPure

MTFa (11.4)a

3

1.41 [0]

Labuz et al. (2021) [20]

OptiSpheric

 + 0.28 µm at 5.15 mm

IsoPure

wOTF (0.51)b

3

1.70 [0.28]

Labuz et al. (2021) [20]

OptiSpheric

 + 0.28 µm at 5.15 mm

ZOE

wOTF (0.51)b

3

2.07 [0.65]

Labuz et al. (2021) [20]

OptiSpheric

 + 0.28 µm at 5.15 mm

Tecnis Eyhance

wOTF (0.51)b

3

2.25 [0.82]

  1. MTFa Area under the modulation transfer function, wOTF Weighted optical transfer function
  2. aEquivalent to 0.2 logMAR according to Vega et al. [29]
  3. bEquivalent to 0.2 logMAR according to Alarcon et al. [30]