Author (Year) | Measurement device | Spherical aberration of cornea | Enhanced Monofocal | Conditions | Aperture (mm) | MTF 50lp/mm |
---|---|---|---|---|---|---|
Borkenstein et al. [21] | OptiSpheric | + 0.28 µm at 5.15 mm | Acunex Quantum | Centered | 3 / 4.5 | 0.61 / 0.40 |
Borkenstein et al. [21] | OptiSpheric | + 0.28 µm at 5.15 mm | Acunex Quantum | 1 mm Decentered | 3 / 4.5 | 0.41 / 0.26 |
Borkenstein et al. [21] | OptiSpheric | + 0.28 µm at 5.15 mm | Acunex Quantum | 5° Tilted | 3 / 4.5 | 0.55 / 0.34 |
Borkenstein et al. [22] | OptiSpheric | + 0.28 µm at 5.15 mm | Lentis Quantum | Centered | 3 / 4.5 | 0.61 / 0.33 |
Borkenstein et al. [22] | OptiSpheric | + 0.28 µm at 5.15 mm | Lentis Quantum | 1 mm Decentered | 3 / 4.5 | 0.49 / 0.26 |
Borkenstein et al. [22] | OptiSpheric | + 0.28 µm at 5.15 mm | Lentis Quantum | 5° Tilted | 3 / 4.5 | 0.58 / 0.35 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | Tecnis Eyhance | Centered | 3 / 4.5 | 0.61 / 0.44 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | Tecnis Eyhance | 1 mm Decentered | 3 / 4.5 | 0.40 / 0.18 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | Tecnis Eyhance | 5° Tilted | 3 / 4.5 | 0.38 / 0.15 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | RayOne EMV | Centered | 3 / 4.5 | 0.41 / 0.22 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | RayOne EMV | 1 mm Decentered | 3 / 4.5 | 0.42 / 0.18 |
Schmid et al. [23] | OptiSpheric | + 0.28 µm at 5.15 mm | RayOne EMV | 5° Tilted | 3 / 4.5 | 0.40 / 0.18 |