Table 1 Data distribution of YD-SX-A, Topcon KR8800 and CR in different groups [D, M (P25, P75)]

From: Comparison of the new self-contained darkroom refractive screener versus table-top autorefractor and cycloplegia retinoscopy in detecting refractive error

Method

Diopter of all participants (1000 eyes)

Diopter of myopic group (783 eyes)

SE(D)

DS (D)

DC(D)

SE(D)

DS (D)

DC(D)

YD-SX-A

-2.06(-3.22, -1.00)

P < 0.001*

-1.50(-2.75, -0.50)

P < 0.001*

-0.50(-1.00, -0.25)

P = 0.032*

-2.50(-3.38, -1.50)

P = 0.729

-2.25(-3.00, -1.25)

P = 0.02*

-0.50(-0.75, -0.25)

P = 0.002*

Topcon KR8800

-2.38(-3.75, -1.25)

P < 0.001*

-2.00(-3.25, -0.75)

P < 0.001*

-0.50(-1.25, -0.25)

P < 0.001*

-2.88(-4.13, -1.88)

P < 0.001*

-2.50(-3.75, -1.50)

P < 0.001*

-0.50(-1.00, -0.25)

P < 0.001*

CR

-1.80(-3.30, -0.80)

-1.50(-2.75, -0.50)

-0.50(-1.00, 0.00)

-2.30(-3.50, -1.50)

-2.00(-3.25, -1.25)

-0.50(-0.75, 0.00)

Method

Diopter of hyperopia group (45 eyes)

Diopter of astigmatism group (289 eyes)

SE(D)

DS (D)

DC(D)

SE(D)

DS (D)

DC(D)

YD-SX-A

0.25(-0.38, 0.94)

P < 0.001*

0.75(0.25, 1.50)

P < 0.001*

-0.75(-1.50, -0.50)

P = 0.159

-1.63(-3.38, -0.69)

P < 0.001*

-0.50(-2.37, 0.37)

P < 0.001*

-2.00(-2.75, -1.37)

P < 0.001*

Topcon KR8800

0.75(-0.06, 2.13)

P < 0.001*

1.50(0.75, 3.00)

P < 0.001*

-1.25(-2.12, -0.50)

P < 0.001*

-1.75(-3.94, 0.38)

P < 0.001*

-0.25(-2.87, 1.12)

P < 0.001*

-2.50(-3.25, -2.00)

P < 0.001*

CR

1.90(1.10, 3.45)

2.50(1.75, 4.12)

-1.00(-2.00, -0.25)

-1.10(-3.60, 0.45)

0.50(-2.25, 2.00)

-2.50(-3.25, -2.00)

  1. *Comparison with CR: P < 0.05