Skip to main content

Table 3 AUC, sensitivity and specificity to detect refractive error with YD-SX-A and Topcon KR8800cutoff values derived from ROC curves

From: Comparison of the new self-contained darkroom refractive screener versus table-top autorefractor and cycloplegia retinoscopy in detecting refractive error

WHO criteria

Myopia(SE < -0.5D)

Hyperopia(SE >  + 2.0D)

Astigmatism(DC < -0.75D)

YD-SX-A

 AUC

0.966

0.974

0.863

 Cutoff

-1.063

-0.562

-0.725

 Sensitivity

90.17%

97.78%

84.08%

 Specificity

90.32%

87.88%

74.26%

Topcon KR8800

 AUC

0.976

0.995

0.980

 Cutoff

-1.188

0.312

-0.875

 Sensitivity

95.15%

97.78%

97.58%

 Specificity

91.24%

96.55%

93.25%