Table 1 Comparison of keratometry of anterior and posterior corneal surfaces between Sirius (Costruzione Strumenti Oftalmici, Italy) and Scansys (MediWorks, Shanghai, China)

From: A comparison of Scansys and Sirius tomography in healthy eyes

 

Device

Minimum

Maximum

Mean ± SD

Correlation

ICC

Difference of mean ± SD*

95% CI of

differences of means

P-value**

r

P-value

Lower

Upper

Anterior

Flat K (D)

Sirius

38.50

46.15

43.18 ± 1.59

0.979

< 0.001

0.985

0.14 ± 0.36

0.02

0.25

0.026

Scansys

39.92

45.85

43.04 ± 1.40

Steep K (D)

Sirius

41.24

47.59

44.35 ± 1.48

0.964

< 0.001

0.980

0.09 ± 0.39

-0.04

0.22

0.156

Scansys

40.78

47.53

44.26 ± 1.36

Astigmatism

J0

Sirius

-0.61

2.23

0.48 ± 0.58

0.957

< 0.001

0.978

0.00 ± 0.17

-0.06

0.05

0.965

Scansys

-0.58

2.27

0.48 ± 0.56

J45

Sirius

-0.32

0.41

0.04 ± 0.18

0.855

< 0.001

0.902

-0.04 ± 0.13

-0.08

0.00

0.069

Scansys

-0.41

0.63

0.08 ± 0.24

Posterior

Flat K (D)

Sirius

-5.67

-6.47

-6.06 ± 0.19

0.956

< 0.001

0.977

-0.14 ± 0.06

-0.16

-0.12

< 0.001

Scansys

-6.28

-5.59

-5.92 ± 0.20

Steep K (D)

Sirius

-6.05

-6.97

-6.46 ± 0.22

0.971

< 0.001

0.985

-0.21 ± 0.05

-0.22

-0.19

< 0.001

Scansys

-6.73

-5.84

-6.25 ± 0.22

Astigmatism

J0

Sirius

-0.45

0.05

-0.19 ± 0.10

0.973

< 0.001

0.984

-0.03 ± 0.02

-0.04

-0.02

< 0.001

Scansys

-0.42

0.02

-0.16 ± 0.09

J45

Sirius

-0.14

0.07

-0.03 ± 0.05

0.578

< 0.001

0.732

-0.03 ± 0.04

-0.05

-0.02

< 0.001

Scansys

-0.08

0.12

0.01 ± 0.04

  1. *Differences of the mean values calculated by subtracting measurements of Sirius from Scansys
  2. **Based on independent samples t-test
  3. D: diopter, N: Number, ICC: intraclass correlation coefficient