Table 2 Comparison of pachymetric indexes, ACD, ACA, IOP sum, HWTW, corneal volume and chamber volume between Sirius (Costruzione Strumenti Oftalmici, Italy) vs. Scansys (MediWorks, Shanghai, China)

From: A comparison of Scansys and Sirius tomography in healthy eyes

 

Device

N

Minimum

Maximum

Mean ± SD

Correlation

ICC

Difference of mean ± SD*

95% CI of

differences of means

P-value**

r

P-value

Lower

Upper

TCT (µm)

Sirius

38

454.00

590.00

529.68 ± 24.14

0.970

< 0.001

0.985

-4.82 ± 5.89

-6.75

-2.88

0.000

Scansys

38

459.00

590.00

534.50 ± 24.08

CCT (µm)

Sirius

38

482.00

613.00

549.89 ± 29.85

0.675

< 0.001

0.794

8.34 ± 22.33

1.00

15.68

0.027

Scansys

38

468.00

593.00

541.55 ± 23.82

ACD (mm)

Sirius

38

2.42

3.62

3.22 ± 0.22

0.957

< 0.001

0.978

0.02 ± 0.07

0.00

0.04

0.111

Scansys

38

2.39

3.61

3.21 ± 0.22

ACA (degree)

Sirius

38

33.00

55.00

46.92 ± 4.24

0.688

< 0.001

0.728

5.26 ± 3.14

4.23

6.30

< 0.001

Scansys

38

33.00

45.00

41.66 ± 2.27

HWTW (mm)

Sirius

38

11.24

12.86

12.17 ± 0.37

0.821

< 0.001

0.901

0.11 ± 0.22

0.04

0.18

0.003

Scansys

38

11.06

12.70

12.06 ± 0.35

Corneal volume (mm3)

Sirius

38

50.40

64.30

57.82 ± 2.48

0.881

< 0.001

0.937

-3.19 ± 1.22

-3.59

-2.78

< 0.001

Scansys

38

53.71

67.91

61.01 ± 2.54

Chamber volume (mm3)

Sirius

38

102.00

236.00

185.71 ± 23.85

0.795

< 0.001

0.884

20.47 ± 16.03

15.20

25.74

< 0.001

Scansys

38

93.75

220.97

165.24 ± 25.82

  1. *Differences of the mean values calculated by subtracting measurements of Sirius HR from Scansys
  2. **Based on independent samples t-test
  3. N: Number, ICC: intraclass correlation coefficient, CCT: central corneal thickness, TCT: thinnest corneal thickness, ACD: anterior chamber depth, ACA: anterior chamber angle, HWTW: horizontal white-to-white