Skip to main content

Table 2 BCDVA and the OQAS optical metrics at 6th months postoperatively among the 4 groups

From: Comparatively analysing the postoperative optical performance of different intraocular lenses: a prospective observational study

 

Mon

SegRef

Dif

EDoF

K/F

P

power

BCDVAa

0.10 (0.10)

0.10 (0.16)

0.10 (0.21)

0.10 (0.10)

7.439

0.059

70.3%

PVA 100%a

-0.04 (0.21)

0.22 (0.2)

0.10 (0.38)

0 (0.16)

30.962

< 0.001***

99.9%

PVA 20%a

0.15 (0.30)

0.40 (0.18)

0.30 (0.42)

0.15 (0.13)

24.910

< 0.001***

99.9%

PVA 9%a

0.40 (0.22)

0.52 (0.18)

0.52 (0.40)

0.40 (0.22)

17.537

0.001***

99.8%

SRa

0.17 (0.05)

0.12 (0.04)

0.13 (0.10)

0.15 (0.04)

18.812

< 0.001***

99.9%

OSIa

1.10 (0.60)

1.90 (1.10)

1.70 (1.20)

1.25 (0.70)

32.257

< 0.001***

99.9%

MTF cut-off frequency (c/deg)b

33.38 (9.49)

18.9 (5.29)

25.97 (11.82)

32.47 (7.85)

13.782

< 0.001***

99.9%

  1. aNonnormally distributed continuous data were described as the median (interquartile range)
  2. bNormally distributed continuous data were described by mean (standard deviation)
  3. Mon = monofocal IOLs; SegRef = segmental refractive IOLs; Dif = diffractive IOLs; EDoF = extended depth of focus IOLs; BCDVA = best corrected distance visual acuity; PVA 100% = predicted visual acuity at contrast of 100%; PVA 20% = predicted visual acuity at contrast of 20%; PVA 9% = predicted visual acuity at contrast of 9%; SR = strehl ratio; MTF = Modulation transfer function; OSI = objective scattering index
  4. The OQAS optical metrics include PVA 100%, PVA 20%, PVA 9%, SR, MTF cutoff frequency, and OSI
  5. BCDVA and PVA were shown in logMAR (logarithm of the minimum angle of resolution)
  6. *P < 0.05, **P ≤ 0.01; ***P ≤ 0.001