Skip to main content

Table 4 Agreement between MS-39 and CASIA SS-1000 in biometric measurements

From: Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT

Parameter

Mean difference ± SD

P

95% LoA

Kf (D)

-0.029 ± 0.260

0.086

-0.54 to 0.48

Ks (D)

-0.007 ± 0.330

0.749

-0.65 to 0.64

Km (D)

-0.024 ± 0.221

0.102

-0.46 to 0.41

TCT (µm)

-4.893 ± 3.331

< 0.0001

-11.4 to 1.6

CCT (µm)

-4.001 ± 3.385

< 0.0001

-10.6 to 2.6

  1. SD = standard deviation; LoA = limits of agreement; Kf = central corneal radius of the flat meridian; Ks = central corneal radius of the steep meridian; Km = mean central corneal radius; TCT = thinnest corneal thickness; CCT = central corneal thickness